7.4.1.1.3 Calibration Valve Assignment
Setup..
Calibration..
Valve assignment..
Valve assignment
Zerogas:
Spangas:
Note!
Entries in this lines are not checked to be pro-
per: The user has to ensure not to configure
a valve that has not been installed!
Emerson Process Management GmbH & Co. OHG
7.4.1 Preparing Calibrations
For each channel a valve has to be assigned
zero gas valve or span gas valve, whereat the
valves can be freely assigned to any channel.
This includes:
Depending on the gases used, this may allow
higher calibration performance.
To do so, enter the SETUP - CALIBRATION -
VALVE ASSIGNMENT menu:
Multi-channel unit:
Select the component to be set in the SELECT
COMPONENT menu.
Configure the valves to be used for the selec-
ted channel.
V1
Available options:
V2
V1 ... V8
Multi-channel unit:
Press the  key to enter the SELECT COM-
PONENT menu to change the settings for a
different channel.
X-STREAM X2
selecting the same combination for all
channels
selecting combinations where one valve
has the same function for several chan-
nels
selecting combinations where one valve
has different functions for several chan-
nels, e.g. the channel 1 zero valve is the
channel 2 span valve.