Instruction Manual
HASX2E-IM-HS
10/2012
6.2.3.2.3.1
6.2.3.2.3.1
Setup..
Calibration..
Interval time..
AutoCal in..
ZeroAll:
ZeroAll:
ZSCalAll:
ZSCalAll:
Emerson Process Management GmbH & Co. OHG
6.2 Menu system - Calibration setup
AutoCal in menu
1 h
15 min
1 h
15 min
Note!
This menu appears only, if at least one interval
time has been set.
The first two lines allow to enter a time to
elapse from the current moment, before the
next zero calibration for all channels is started.
Preset is the remaining time until the next zero
calibration starts.
Possible values: Any value up to the preset
time interval.
Lines 3 & 4 allow to enter similar data for a
combined zero & span calibration.
Possible values: Any value up to the preset
time interval.
Note!
In case there is an interval time specified
for one procedure only (ZeroAll or ZSCalAll;
6.2.3.2.3, page 6-26), the related lines
for the other procedure will be hidden in this
menu!.
X-STREAM X2
6-27