7.5.3.4 Finalizing the Sensor Replacement
Disconnect the analyzer from power
Re-install the sensor unit into the analyzer
Close the housing. Take care to use all
screws, especially if the instrument is to
be used in hazardous areas!
or
close the rear panel cover and secure it
with the screw nut.
Special conditions and instructions for start-up after maintenance apply
to instruments to be operated in hazardous areas!
Not observing these conditions and instructions may cause explosions!
See the associated manuals, provided with instruments for use in hazardous
areas, for more information.
In a next step for proper measuring results,
perform a zero and a span calibration at least
for the channel with the replaced sensor.
To ensure proper disposal, send back the old
sensor to the EMERSON Process Manage-
ment factory (or to your local sales office) or
to an industrial waste management contractor
for waste disposal.
Emerson Process Management GmbH & Co. OHG
7.5 Replacing Worn Out Sensors
EXPLOSION HAZARD
X-STREAM X2
7-53