X-STREAM X2
On user request restore and
overwrite current configurati-
(SvcPort --> CfgData,
7.6.4.3, page 7-59 )
Overwrites after factory analy-
zer startup, or on user request
(CfgData --> UserData,
7.6.1, page 7-53)

Fig. 7-17: Relationship of Supported Data Sets and Where to Find Further Information

7-58
7.7
Save / Restore Configuration Data Sets
SvcPort
Backup/Restore
to/from
external device
Backup on user request
on in RAM
(CfgData --> SvcPort ,
CfgData
configuration
used during
operation
UserData
configuration
saved to
FRAM
Note!
Before using the SvcPort read the
instructions in
7.6.4.2, page 7-57 )
Overwrites on user request
(FactData --> CfgData,
7.6.3, page 7-55)
Overwrites after factory
analyzer startup
Automatically overwrites in case
of faulty checksum, or
on user request
(UserData --> CfgData,
7.6.2, page 7-54)
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASX2E-IM-HS
7.6.4.1, page 7-56.
FactData
Factory setup
configuration
10/2012