Instruction Manual HASAxE-IM-HS 07/2006 Gas Analyzer Series Instruction Manual www.EmersonProcess.com...
Page 2
• Read all instructions prior to installing, operating, and servicing the product. • If you do not understand any of the instructions, contact your Emerson Process Management (Rosemount Analytical) representative for clarification. • Follow all warnings, cautions, and instructions marked on and supplied with the product.
Highlights an operation or maintenance procedure, practice, condition, state- ment, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness. NOTE Highlights an essential operating procedure, condition or statement. Emerson Process Management GmbH & Co. OHG...
Electrostatic discharge (ESD) Explosion Hazard! Heavy Instrument! Toxic! Harmful (to Health)! Disconnect from Mains! All graphical symbols used in this product are from one or more of the following standards: EN61010-1, IEC417, and ISO3864. Emerson Process Management GmbH & Co. OHG...
Page 5
Use only fuses of the correct type and current ratings as replacements. Using repaired fuses and short circuiting of fuse holders is prohibited. Emerson Process Management GmbH & Co. OHG...
Page 6
Aviod inhalation of the exhaust gases at the exhaust fitting. Connect exhaust outlet to a safe vent. Check vent line and connections for leakage. Keep all fittings tight to avoid leaks. See section 7-2, page 7-2 for leak test instructions. Emerson Process Management GmbH & Co. OHG...
Page 7
Take care the wall or stand the unit is intended to be installed at is solid and stable to hold the units! HIGH TEMPERATURES ! While working with photometers and/or thermostated components inside the analyzers hot components may be accessible! Emerson Process Management GmbH & Co. OHG...
Page 8
Before opening gas paths they must be purged with ambient air or neutral gas (N2) to avoid hazards caused by toxic, flammable, explosive or harmful to health sample gas components! Emerson Process Management GmbH & Co. OHG...
Page 9
If double insulation is not available signal cables inside the analyzer must be installed in a way that a distance of at least 5 mm is ensured permanently (e.g. by utilizing cable ties). Emerson Process Management GmbH & Co. OHG...
Page 10
Any interruption of the earth connector inside or outside the instrument or disconnecting the earth terminal may cause potential electrical shock hazzard! Intended interruption of protective earth connections is not permitted! Emerson Process Management GmbH & Co. OHG...
To stay in compliance with regulations regarding electromagnetic compatibility it is recommended to use only shielded cables, as optionally available from Emerson Process Management or equivalent. Customer has to take care that the shield is connected in proper way ( section 4-5, page 4-29).
For air transportation of magnetic material the IATA instructions have to be observed: Magnetic fields are not allowed to penetrate the package, if necessary the magnets have to be shorted using a metal plate. S-10 Emerson Process Management GmbH & Co. OHG...
Page 13
Software menu structure, how to navigate and menu entries descriptions ......5 startup procedures, checking the instrument‘s setup ..... 6 Basic procedures ..........7 Maintenance procedures ......... 7 Status messages and troubleshooting ..8 Modbus Parameters .......... 9 Emerson Process Management GmbH & Co. OHG...
Digital Outputs ..................... 1-18 1-8-2 Digital Inputs ......................1-18 1-8-2-1 Remotely Activating a Calibration Procedure using IN 1 to IN 3 ....1-19 1-8-2-2 Valve Control....................1-20 1-8-2-3 Remote Pump Control ..................1-20 Emerson Process Management GmbH & Co. OHG...
Page 15
Hints on Wiring Signal Inputs and Outputs ............4-29 4-5-1 Electrical Connections in General..............4-29 4-5-2 Wiring Inductive Loads ..................4-31 4-5-3 Driving Multiple Loads ..................4-31 4-5-4 Driving High Current Loads................4-32 Emerson Process Management GmbH & Co. OHG...
Page 16
Signal Damping Setup Menu ..............5-30 5-4-3-3-2 Cross Interference Setup Menu ..............5-31 5-4-3-3-3 AD Conversion Setup Menu ..............5-32 5-4-3-4 In/Outputs Setup Menu ................... 5-33 5-4-3-4-1 Analog Output Setup Menu ................ 5-34 Emerson Process Management GmbH & Co. OHG...
Page 17
Display Information Setup ..................6-4 6-2-3 Calibration Data Setup ..................6-4 6-2-4 Analog Output Setup ..................... 6-7 6-2-5 Concentration Alarms Setup ................6-12 Chapter 7 Maintenance Abstract ........................7-1 Performing a Leak Test....................7-2 Emerson Process Management GmbH & Co. OHG...
Page 18
Restore UserData to CfgData ................7-47 7-6-3 Restore FactData to CfgData ................7-48 7-6-4 Save / Restore to an External Device..............7-49 7-6-4-1 Save CfgData to COMPort ................7-50 7-6-4-2 Restore COMPort to CfgData ................. 7-52 Emerson Process Management GmbH & Co. OHG...
Page 19
Chapter 10 Service Information 10-1 10-1 Return of Material ..................... 10-1 10-2 Customer Service ....................10-2 10-3 Training ........................10-2 Appendix Modbus Implementation ................... A-2 EC Declaration of Conformity ................A-12 Block Diagrams ....................... A-13 Emerson Process Management GmbH & Co. OHG...
(e.g. intrinsically safe conditions for analyzers in hazardous areas, cells). related to installation, operation, maintenance etc. For such applications refer to the separate instruction manuals, delivered together with the analyzers. Emerson Process Management GmbH & Co. OHG...
Page 24
Facilitates compensation of atmospheric pressure variations to improve precision of results ( measurement specifi- cations, page 3-17). Special sensors for e.g. corrosive gas on request. • Flow measurement A flow sensor (option) can be used to Emerson Process Management GmbH & Co. OHG...
The following sections 1-2 to 1-3 give detailed their appearance. descriptions for all available configurations. Table top and rack mount version (see section 1-2, page 1-5) Field housing (see section 1-3, page 1-7) Emerson Process Management GmbH & Co. OHG...
4x20 characters alphanumeric display "Home" key LED (red) "Enter" key LED (red) 4 keys for editing and entering menus LED (green) Fig. 1-1: X-STREAM GP, GPS, front side view Emerson Process Management GmbH & Co. OHG...
Installation, startup and maintenance are described in detail in a separate instruction manual, shipped together with each such analyzer, and are not subject of this current instruction manual! Emerson Process Management GmbH & Co. OHG...
25 kg (57 lbs), depending on installed options! The upper part of the front door is designed to work as a carrying handle, see fig. 1-4. Use two persons or a suitable lifting device to move or carry the instrument! Emerson Process Management GmbH & Co. OHG...
1-3 X-STREAM F Cable glands for mains and signal cables Gas inlet and outlet fittings and purge gas outlet fitting Purge gas inlet fitting 4 supports for wall mounting Fig. 1-6: X-STREAM F, bottom view Emerson Process Management GmbH & Co. OHG...
X-STREAM F Terminals for signal cables Mains EMI filter Cable glands for mains and signal cables Mains terminals with integrated fuses Fig. 1-7: X-STREAM F, mains and signal terminals (front door removed) 1-10 Emerson Process Management GmbH & Co. OHG...
6/4 mm or 1/4") or other fitting materials (on request) may be used. 1-5-4 Piping Analyzers are piped with Viton or PTFE (ø 6/4 mm). Other materials (e.g. stainless steel) are used optionally, depending on application. 1-12 Emerson Process Management GmbH & Co. OHG...
(IN K1) Fig. 1-8: Single channel or serial tubing Gas outlet (OUT K1) Analyzer Gas outlet (OUT K2) Gas inlet (IN K1) Gas inlet (IN K2) Fig. 1-9: Dual channel parallel tubing Emerson Process Management GmbH & Co. OHG 1-13...
Internal or Int+Ext ( 5-4-3-5, page 5-41). Internal or Use ch2 ( 5-4-3-5, page 5-41). Valve control is supported by a corresponding menu line, by selecting the autocal mode or remotely by digital inputs. Emerson Process Management GmbH & Co. OHG 1-15...
Refer to the technical data section of your maximum load of 30 V / 1 A / 30 W each! analyzer variation for a more detailled infor- mation about relay status signals ( 2-2, page 2-5 et sqq.)! Emerson Process Management GmbH & Co. OHG 1-17...
Input resistance: 57.5 kΩ (Voltages to be measured against the reference terminal labeled "IN GND"). Inputs protected against overvoltage up to approx. 40 V. Open inputs are at LOW level. 1-18 Emerson Process Management GmbH & Co. OHG...
Inputs E and F start at the same time: F is rejected because of the lower priority in the sequence IN 1 - IN 2 - IN 3. Fig. 1-14: Digital Inputs - Priorities Emerson Process Management GmbH & Co. OHG 1-19...
Action / Input Open valve V4 Open valve V1 Open valve V2 Open sample gas valve Close all valves H: HIGH L: LOW X: don't care Table 1-1: Digital Inputs IN4-IN7, array 1-20 Emerson Process Management GmbH & Co. OHG...
+/- 10 % of the nominal supply voltage! Input voltage range 85 - 264 V , 47 - 63 Hz Rated input current standard 0,75 - 0,35 A max. with thermostated physics 2 - 1 A max. Emerson Process Management GmbH & Co. OHG...
Page 45
8 digital outputs 2 thresholds per channel, (optically isolated, common ground) sample gas valve, zero gas valve V4, span gas valve V1, span gas valve V2 “Open Collector”, max. 30 V / 30 mA Emerson Process Management GmbH & Co. OHG...
Page 46
Purge medium (e.g. for minimizing CO inter- For pressure and flow specification consult ference, safety in events of internal leakage of factory or your local EMERSON Process Ma- flammable or/and aggressive gas etc.) must nagement sales office! be dry, clean and free of corrosive and solvent components and on ambient temperature (at least 68...95 F / 20...35 °C) !
Instruction Manual HASAxE-IM-HS X-STREAM 05/2006 2-2 Model Specific Technical Data 2-2 Model Specific Technical Data 2-2-1 X-STREAM GP, GPS Tabletop or Rack Mount Version Fig. 2-1: X-STREAM GP, GPS - dimensions [approx. mm (inch)] Emerson Process Management GmbH & Co. OHG...
), using conductor sleeves is not required Cable entry through cover via 1 strain relief Power fuses Two fuses are located nearby the power terminals. Fuse ratings: AC 230 V, 3.15 A, 5x20 mm Emerson Process Management GmbH & Co. OHG...
All signal lines need to be connected to screw terminals located at the analyzer‘s rear panel. Available signals: standard: Analog signal outputs Relay status signals Modbus interface (RS232; RS 485) optional: Digital inputs/outputs Detailed pin assignment 4-4 Installation, page 4-6. Emerson Process Management GmbH & Co. OHG...
Two fuses are located in the power appliance. Fuse ratings: AC 230 V, 3.15 A, 5x20 mm Power appliance with EMI filter Fuse holder Signal sockets/plugs Fig. 2-3: X-STREAM GPS, (sockets version) - power and signals connectors Emerson Process Management GmbH & Co. OHG...
Page 51
25 pole submin-D-plugs and sockets, located at the analyzer‘s rear panel (fig. 2-3). Available signals: standard: Analog signal outputs Relay status signals Modbus interface (RS232; RS 485) optional: Digital inputs/outputs Detailed pin assignment 4-4 Installation, page 4-6. Emerson Process Management GmbH & Co. OHG...
IP 66 (EN 60529) / NEMA 4X for outdoor installation Analyzer must not be exposed to direct sun light Gas fittings: quantity: max. 8 specification: 6/4 mm PVDF optional 6/4 mm or 1/4", stainless steel, other on request 2-10 Emerson Process Management GmbH & Co. OHG...
Page 53
AC 230 V, 3.15 A, 5x20 mm Power terminals with integrated Power cable gland fuse holders EMI power filter Protective earth terminal (PE) Fig. 2-5: X-STREAM F - power terminals / fuse holders Emerson Process Management GmbH & Co. OHG 2-11...
400 and 1000 ppm, ANALOG OUTPUT SIGNAL SCALING menu, page 5-35 ). The measuring ranges are also shown in the INFO - RANGES.. menu, page 5-57. Emerson Process Management GmbH & Co. OHG 2-13...
55 °C or built into a thermostatted box, covering all physical components and completely thermostatted. Section 3-1-2 covers both IR and UV measure- ment. Emerson Process Management GmbH & Co. OHG...
As result the detector gives an alternating voltage signal which is passed to electronics for further processing. Emerson Process Management GmbH & Co. OHG...
The flow detector signal is passed to elec- tronics conditioning the signal and converting it into a usefull format. Emerson Process Management GmbH & Co. OHG...
Analysis cell, reference side Filter cell Preamplifier Filter cell Pyroelectrical detector (solid state detector) Gas detector Preamplifier Chopper wheel Fig. 3-3: IR Photometer Assembly Principle, left side: with Gas Detector, right side: with Pyro Electrical Detector Emerson Process Management GmbH & Co. OHG...
Page 62
The chopper windows and those at the filter cell remain as only possibly polluted windows. All of them are accessible once the analysis cell is removed. Emerson Process Management GmbH & Co. OHG...
The sensor‘s designed life under constant conditions of 20 °C is approx. 900,000 hrs. The lifetime at 21 % oxygen is then calculated to approx. 42,857 hrs, corresponding to approx. 5 years. 3-12 Emerson Process Management GmbH & Co. OHG...
Supplying cells continuously with dry sample gas of low Table 3-4: Electrochemical Oxygen Measurement, grade oxygen concentration or with sample cross interference by accompanying gases Emerson Process Management GmbH & Co. OHG 3-13...
Depending on application it is possible to supply a reference gas to the bridge‘s reference side. The output signal in this case is proporti- onal to the difference of the thermal conductivities of sample and reference gas. 3-14 Emerson Process Management GmbH & Co. OHG...
Sensor Internnal gas path Sample gas inlet and output Sample gas inlet and output Reference side inlet and output PT 100 sensors Metal block Metal block Heater for thermostatting Emerson Process Management GmbH & Co. OHG 3-15...
0 - 5 % Other components and configurations on request Dew point below ambient temperature Higher concentrations decrease sensor lifetime Specification for lowest range to be verified Table 3-6: Gas Components and Measuring Ranges, examples 3-16 Emerson Process Management GmbH & Co. OHG...
Constant pressure and temperature Starting from 20°C/68 °F (to 0 °C / 32 °F and Dependent on integrated photometer bench to +50 °C / 122 °F) Paramagnetical oxygen measurement (pO Table 3-7: Measurement Performance Specifications Emerson Process Management GmbH & Co. OHG 3-17...
Take care to meet the permissible least sheltering against rain is recommended. ambient temperatures as given in the techni- cal data section! Instruments must not be exposed to direct sunlight nor sources of heat. Emerson Process Management GmbH & Co. OHG...
77 °F (25 °C). Pressure and flow must be within the limits given by the measurement specifications table 3-7 3-4, page 3-17). Emerson Process Management GmbH & Co. OHG...
When it is necessary to open gas paths seal the analyzer‘s gas fittings using PVC caps to Fig. 4-0: Example of gas fittings label avoid pollution of the internal gas path by moisture, dust, etc. Emerson Process Management GmbH & Co. OHG...
Fig. 4-1: Bypass Mode Installation Internal Solenoid Valve Block Supply overpressure for all gases is limited to 0.7 to 7.2 psi (50 to 500 hPa) when the analyzer is equipped with an internal solenoid valve block. Emerson Process Management GmbH & Co. OHG...
The analyzers do not provide a power switch and are operable when connec- ted to power. Installation instructions for X-STREAM GP, X-STREAM GPS Page 4-6 X-STREAM F Page 4-20 Hints on wiring signal inputs and outputs Page 4-28 Emerson Process Management GmbH & Co. OHG...
(fig. 4-2). when connected to power. Fixing holes for rack mounting (rack mount version only) (0.3 x 0.4 inches / 7.5 x 10.5 mm) Fig. 4-2:Frontal view Emerson Process Management GmbH & Co. OHG...
Out2, ...). This avoids confusion during re-installation when the analyzer had to be disconnected for whatever reason. Gas inlets and outlets Quantity: max. 8 (+ 1 optional purge gas fitting) Specification: 6/4 mm PVDF optional 6/4 mm or 1/4", stainles steel, other on request Emerson Process Management GmbH & Co. OHG...
5 / (+) 4 (0) - 20 mA, channel 2 ≤ 500 Ω Burden: R Legend: Pin # / Signal Fig. 4-4: Socket X1 - pin assignment Note! Take care of the special installation instructions in section 4-5! Emerson Process Management GmbH & Co. OHG...
Page 86
The power power inlet appliance is of type IEC. Use a standard power cable with IEC connector complying with your local requirements to supply the instrument with power. PE = Protective Earth Fig. 4-9: IEC power input plug 4-12 Emerson Process Management GmbH & Co. OHG...
A label fixed to the inner side of the terminals cover shows how the terminals are assigned. Gas inlets and outlets Quantity: max. 8 (+ 1 optional purge gas fitting) Specification: 6/4 mm PVDF optional 6/4 mm or 1/4", stainles steel, other on request Emerson Process Management GmbH & Co. OHG 4-13...
Common Common Common RXD0 not used RXD1 not used TXD1 not used Common TXD0 Fig. 4-12: Modbus Interface Terminals Note 2! X-STREAM analyzers are to be considered a DTE (Data Terminal Equipment). Emerson Process Management GmbH & Co. OHG 4-15...
RS and relay contact signals). Failure COM Failure NO Maintenance, off-spec COM Maintenance, off-spec NO Function check COM Function check NO Fig. 4-14: Relay Status Terminals 4-16 Emerson Process Management GmbH & Co. OHG...
Terminal 13 Open V4 Terminal 14 Open V1 Terminal 15 Open V2 Terminal 16 Open sample valve or close all Terminal 17 DIG In GND Fig. 4-15: Digital Input & Output Terminals Emerson Process Management GmbH & Co. OHG 4-17...
Page 92
Line (L) Protective Earth (PE) Neutral (N) Fig. 4-16: Power terminals 4-18 Emerson Process Management GmbH & Co. OHG...
Page 93
• associated openings of the cover and finally attach the terminals cover to the instru- • ment's rear panel utilizing all 4 torx screws. Emerson Process Management GmbH & Co. OHG 4-19...
Take care to use anchors and bolts specified to be used for the weight of the units! Take care the wall or stand the unit is intended to be installed at is solid and stable to hold the units! 4-20 Emerson Process Management GmbH & Co. OHG...
Gas inlet and outlet fittings 4 cable glands for power and signal cables Purge gas inlet fitting (option) Power terminals with integrated fuses Fig. 4-18: X-STREAM F -Allocation of terminals and gas fittings Emerson Process Management GmbH & Co. OHG 4-21...
Page 96
Uncover the shielding Feed cable through gland nut and into fixing element Put the shielding net over the element the way that it covers the o-ring 2 mm. 4-22 Emerson Process Management GmbH & Co. OHG...
4-22! ≤ 500 Ω Burden: R Fig. 4-19: X-STREAM F - Analog output terminals ELECTRICAL SHOCK HAZARD! Verify power cables are disconnected and/or instrument is de-energized prior to working at the terminals! Emerson Process Management GmbH & Co. OHG 4-23...
RXD0 not used RXD1 not used TXD1 not used Common TXD0 Note 2! X-STREAM analyzers are to be considered a DTE (Data Terminal Equipment). Fig. 4-20: X-STREAM F - Modbus interface terminals 4-24 Emerson Process Management GmbH & Co. OHG...
Terminal Signal Failure COM Failure NO Maintenance, off-spec COM Maintenance, off-spec NO Function check COM Function check NO Fig. 4-22: X-STREAM F - Relay Status Terminals Emerson Process Management GmbH & Co. OHG 4-25...
4-5 and of the cable gland strip is reserved for the digital inputs and assembly instruction on page 4-22! outputs. Digital Inputs DigitalOutputs Fig. 4-23: Digital Input & Output Terminals 4-26 Emerson Process Management GmbH & Co. OHG...
When done, fix the power cable by tightening the outer cable gland nut. L= Line N=Neutral PE=Protective Earth Location Power cable gland for label Fig. 4-24: Power terminals Emerson Process Management GmbH & Co. OHG 4-27...
Page 102
Ensure the earthing conductor is connected! After all connections are established in correct manner and verified, close the front door and secure it utilizing • the two sash fasteners. 4-28 Emerson Process Management GmbH & Co. OHG...
It is recommended to use shielded cables for signal lines, only! Shield has to be connected to the housing at both ends of one connection (fig. 4-25). Fig. 4-25: Shielded Signal Cable, shield connected at both ends Emerson Process Management GmbH & Co. OHG 4-29...
This gives an advantage when both instruments are supplied by different supply networks (e.g. when installed in dif- ferent buildings). Fig. 4-27: Double-shielded Signal Cable, shields connected at both sides 4-30 Emerson Process Management GmbH & Co. OHG...
(fig. 4-28). Suitable filter components are available on request for standard valves. Fig. 4-29: ”Serial” Wiring Fig. 4-28: Suppressor Diode for Inductive Loads Emerson Process Management GmbH & Co. OHG 4-31...
Run wires for one load in parallel as long as even practicable Load (drill if possible) External relay Analyzer output Fig. 4-30: Running Supply Lines "Parallel” Fig. 4-31: Driving High Current Loads 4-32 Emerson Process Management GmbH & Co. OHG...
In case of X-STREAM F the display, LEDs and necessary information to operate, calibrate and keys are located behind a protective glass and function check the instrument. the keys (sensors) are operated by a magnetic tool. Fig. 5-1: X-STREAM user interface Emerson Process Management GmbH & Co. OHG...
"FAILURE“. If this red LED is • Flashing: Maintenance Request, function check or out of specification. Off: Measurement OK • The 3rd LED gives power information: Illuminated (green): Power On Dark: Power Off Emerson Process Management GmbH & Co. OHG...
UP / DOWN keys: Mode Function Measure Exit measuring screen Browse - Select a menu line - Open previous/next menu page if placed in a line starting with / Edit increase /decrease the input value Emerson Process Management GmbH & Co. OHG...
UP/DOWN keys the magnets need to be vertical. horizontal versus vertical orientation The HOME/ENTER keys require a horizontal orientation (represented by horizontal ellipses surrounding the key symbols). Fig. 5-2: X-STREAM magnetic tool Emerson Process Management GmbH & Co. OHG...
MEASURING SCREEN flow, temperature nothing (left blank) • By default the upper line shows the channel 1 data. For dual channel instruments the second line gives the data for the second measuring channel. Emerson Process Management GmbH & Co. OHG...
In addition a function check active from editing a parameter automatically is reset when returning to the measuring screen! Emerson Process Management GmbH & Co. OHG...
Page 113
A dual channel analyzer requires to select the channel the measurement related parameters Component: CO2.1 are to be edited. In this case a selection menu (i)Back: Press appears automatically when a choice is Emerson Process Management GmbH & Co. OHG...
Access level Access Code Status 00000001 00000002 00000003 It is recommended to change the access codes when using access levels ( 5-4-3-1-2 Menu Access Setup, page 5-21). Emerson Process Management GmbH & Co. OHG...
-COMMAND EXECUTED- Command execution confirmation: Confirms a procedure (e.g. calibration) has been started. -CANCELLED- Command cancellation confirmation: Confirms a procedure (e.g. calibration) has been cancelled. Emerson Process Management GmbH & Co. OHG...
Access level 2 (expert) Access level 3 (system integrator / administrator) Access level 4 (service level) Menu title Parent menu title xxx.. xxxx and current menu title Optional component selection menu page 5-7) 5-10 Emerson Process Management GmbH & Co. OHG...
Control Menu 5-4-2, page 5-13 Control… Setup… Setup menu 5-4-3, page 5-20 Status… Info… Status menu 5-4-4, page 5-47 MAIN MENU Info menu 5-4-5, page 5-55 Emerson Process Management GmbH & Co. OHG 5-11...
Instruction Manual HASAxE-IM-HS X-STREAM 05/2006 5-4 Menu System - Control Menu Fig. 5-3: X-STREAM Software menu structure 5-12 Emerson Process Management GmbH & Co. OHG...
Pressing the key returns to the optional gas Note! component selection menu to perform the For detailed descriptions of how to perform same action for the other measuring channel. calibrations chapter 7 Maintenance. 5-14 Emerson Process Management GmbH & Co. OHG...
Pressing the key returns to the optional gas For detailed descriptions of how to perform component selection menu to perform the calibrations chapter 7 Maintenance. same action for the other measuring channel. Emerson Process Management GmbH & Co. OHG 5-15...
/ span calibration menus while the 4 line allows to start zero and span calibration by pressing just one key. Note! For detailed descriptions of how to perform calibrations chapter 7 Maintenance. 5-16 Emerson Process Management GmbH & Co. OHG...
This line gives the measured concentration CO.1 13.304 ppm Procedure None Time Information about which procedure is active (none, purging, zero 1, zero 2, span 1, span 2) Remaining time till end of active procedure Emerson Process Management GmbH & Co. OHG 5-17...
(except for None: all valves are closed). Dual channel instrument: Pressing the key returns to the optional gas component selection menu to perform the same action for the other measuring channel. 5-18 Emerson Process Management GmbH & Co. OHG...
This page allows to acknowledge (=reset) all status messages:Just press the key to per- Status! form an acknowledgement. When processed the screen changes temporarily to show a command execution -COMMAND EXECUTED- confirmation. Emerson Process Management GmbH & Co. OHG 5-19...
Select the value to be displayed in the related line of the measuring screen: Comp-1, Temp-1, Pres-1 Line 1: Comp1 Comp-2, Temp-2, Pres-2 Line 2: Comp2 Blank (none) Line 3: Blank Component.. Component setup menu menu page 5-4-3-8, page 5-25 Emerson Process Management GmbH & Co. OHG 5-21...
Page 128
Current temperature sensor 1 value Set pressure unit:: Pa, hPa, mbar, Bar, psi Pressure Unit: DecimalPlaces: Set pressure value's decimal places: Pressure 998.1 hPa 0 to 4 menu page Current pressure value 5-22 Emerson Process Management GmbH & Co. OHG...
Select your prefferred language for the analyzer Language software. Number of and languages may change depending on software revision. Language: English Currently available: English, French, German, Italian, Spanish Emerson Process Management GmbH & Co. OHG 5-23...
Code 1: 00000001 the related access levels. Code 2: 00000002 Figure shows the factory default settings. Code 3: 00000003 We recommend changing the default settings when making menu page use of access codes! 5-24 Emerson Process Management GmbH & Co. OHG...
Input range: 0 .. 600 seconds Interval time setup menu Note! 5-4-3-2-2, page 5-28 For a detailed description of how to perform calibrations chapter 7 Maintenance. 5-26 Emerson Process Management GmbH & Co. OHG...
Dual channel instrument: Pressing the key while any line is selected returns to the optional gas component selec- tion menu to open the same menu for the other measuring channel. Emerson Process Management GmbH & Co. OHG 5-27...
Enter the time between two span calibrations SpanCal: 10 h when in autocal mode. Input range for both values: 0 .. 999 h Note! For detailed descriptions of how to perform calibrations chapter 7 Maintenance. 5-28 Emerson Process Management GmbH & Co. OHG...
"manual" to achieve best measu- AD conversion.. ring results. Signal damping setup menu 5-4-3-3-1, page 5-30 Cross interference setup menu 5-4-3-3-2, page 5-31 AD conversion setup 5-4-3-3-3, page 5-32 Emerson Process Management GmbH & Co. OHG 5-29...
The analyzer's total signal propagation time (commonly t90 time) is the sum of the signal damping time and the physical propagation delay time caused by e.g gas flow and sensor properties. 5-30 Emerson Process Management GmbH & Co. OHG...
Requires using pure gas for span calibration! Note! Setting this parameter to Yes is valid for the next span calibration only and will be reset to No automatically after calibration has finished. Emerson Process Management GmbH & Co. OHG 5-31...
The Comp2Offset offset's unit are counts. MuxOffset Select the last line to setup the AD converters offset for the secondary measurement (This is a multiplexed converter): The offset's unit are counts. 5-32 Emerson Process Management GmbH & Co. OHG...
The line "PumpControl" shows up only if the parameter "Pump" in the INSTALLED OPTIONS setup menu is Yes. The line "Valve assignment" shows up only if the parameter "Valves" in the INSTALLED OPTIONS setup menu other than None. Emerson Process Management GmbH & Co. OHG 5-33...
"Signal range" parameter 5-4-3-4-1-2, page 5-36) *) Select one of these modes only to do a quick check if the analog output level settings are still correct. Table 5-1: Analog Output Signal Selection Emerson Process Management GmbH & Co. OHG 5-35...
0 mA -0.2 mA Dead-Zero similiar HL4-20mA above 4 … 20 mA 3,8 mA 20,5 mA >21 mA 0 mA 3.8 mA Life-Zero Table 5-2: Analog Output Signal Setting & Operation Modes Emerson Process Management GmbH & Co. OHG 5-37...
(here: 1000 ppm) Dual channel instrument: Pressing the key while any line is selected returns to the optional gas component selec- tion menu to open the same menu for the other measuring channel. 5-38 Emerson Process Management GmbH & Co. OHG...
The analog output signal changes accordingly. tion menu to open the same menu for the Repeat editing the parameter until the analog other measuring channel. output creates the expected signal. Emerson Process Management GmbH & Co. OHG 5-39...
• has the same function for both channels selecting combinations where one valve • has different functions for both channels, e.g. the channel 1 zero valve is the channel 2 span valve. 5-40 Emerson Process Management GmbH & Co. OHG...
Select the serial interface baud rate. Baud rate: 19200 Available options: Parity: None 2400, 4800, 9600, 19200 Select if a parity bit is used or not. Available options: menu page None, Even, Odd 5-42 Emerson Process Management GmbH & Co. OHG...
Dual channel instrument: Pressing the key while any line is selected returns to the optional gas component selec- tion menu to open the same menu for the other measuring channel. Emerson Process Management GmbH & Co. OHG 5-43...
ConfigData set with data received via the serial interface. The current configuration data is not over- written! Incoming data is erased after being compared to the RAM data. 5-44 Emerson Process Management GmbH & Co. OHG...
Page 151
If there is a failure, the UserData settings are restored into RAM overwriting the ConfigData. This ensures the instrument remains operable. Chapter 7 for a detailled description of these functions. Emerson Process Management GmbH & Co. OHG 5-45...
Alarms will be activated when the currently measured values exceed the given limits during calibration. Note! This behaviour may cause trouble when the instrument is connected to e.g. a data aqui- sition system. 5-46 Emerson Process Management GmbH & Co. OHG...
Off spec status screen 5-4-4-4, page 5-51 Calibration status menu 5-4-4-5, page 5-52 Calibration.. Measurement status menu Measurement.. 5-4-4-6, page 5-53 Alarms.. menu page Alarms status menu 5-4-4-7, page 5-54 Emerson Process Management GmbH & Co. OHG 5-47...
Optional gas component selection menu - Select the component of interest. ADC-Error Dual channel instrument: Pressing the key returns to the optional gas component selection menu to open the same menu for the other measuring channel. 5-48 Emerson Process Management GmbH & Co. OHG...
Optional gas component selection menu - Select the component of interest. ZeroCalTolC SpanCalTolCh Dual channel instrument: Pressing the key returns to the optional gas component selection menu to open the same menu for the other measuring channel. Emerson Process Management GmbH & Co. OHG 5-49...
Optional gas component selection menu - Select the component of interest. LocalAccess Dual channel instrument: Pressing the key returns to the optional gas component selection menu to open the same menu for the other measuring channel. 5-50 Emerson Process Management GmbH & Co. OHG...
Select the component of interest. Lineariser Normal Temperature SignalRange Dual channel instrument: Pressing the key returns to the optional gas component selection menu to open the same menu for the other measuring channel. Emerson Process Management GmbH & Co. OHG 5-51...
Purging: the instrument is purging the gas path The last line shows the remaining time for the given procedure. Note! For a detailed description of calibration status and how to perform calibrations chapter 7 Maintenance. 5-52 Emerson Process Management GmbH & Co. OHG...
704 mA Shows IR source current. menu page Dual channel instrument: Pressing the key returns to the optional gas component selection menu to open the same menu for the other measuring channel. Emerson Process Management GmbH & Co. OHG 5-53...
(e.g. CO.1 Alarm Level1). Dual channel instrument: Pressing the key returns to the optional gas component selection menu to open the same menu for the other measuring channel. 5-54 Emerson Process Management GmbH & Co. OHG...
Latin America.. world region. Asia Pacific.. menu page Factory settings information menu 5-4-5-2, page 5-57 Factory settings.. Installed options.. Installed options info menu 5-4-5-1, page 5-58 menu page Emerson Process Management GmbH & Co. OHG 5-55...
Dual channel instrument: Pressing the key returns to the optional gas component selection menu to open the same screen for the other measuring channel. 5-56 Emerson Process Management GmbH & Co. OHG...
For this reasons access is pro- tected by access level code 4 and permitted to specially trained personnel only! Access level 4 code must be entered to access this menu. Emerson Process Management GmbH & Co. OHG 5-57...
These pages provide information about Valves Int+Ext installed options. COM-Interf To change any of these status enter the Pump "INSTALLED OPTIONS SETUP" menu ( menu page 5-4-3-5, page 5-41). DigitalInp Pressure Internal Flow monitor menu page 5-58 Emerson Process Management GmbH & Co. OHG...
It is assumed that the operator is familiar how to navigate through menus and how to activate procedures, as described in chapter 5. Furthermore the instrument should be installed according the instructions given in chapter 4. Emerson Process Management GmbH & Co. OHG...
Factory default setting is "no code required". Select your prefferred language for the analyzer Language software. Number of and languages may vary depending on software revision. Language: English Currently available (to be extended) : English, German, French, Spanish, Italien Emerson Process Management GmbH & Co. OHG...
ENTER key this language is set and the menu's language is updated immediately. If the selected language is not the one desired, repeat the sequence of last three keys multiple times until the desired language is set. Emerson Process Management GmbH & Co. OHG...
Take the correct values from your gas supplier's certificate. Proper values only assure ZeroGas: 0.0 ppm faultless measuring results. For multichannel instruments enter the values SpanGas: 500.0 ppm for each channel separately. Emerson Process Management GmbH & Co. OHG...
Page 169
Enter a purge time that assures the measuring cell is sufficiently filled after switching from one to the other gas. Emerson Process Management GmbH & Co. OHG...
Page 170
Factory setting is 2 seconds, values from 2 to 60 seconds are accepted, for multichannel instruments independently for each channel. "Cross interference" settings are available for multichannel instruments only. Single channel instruments don't show this line, opening a next screen: Emerson Process Management GmbH & Co. OHG...
PumpControl: Menu (switched ON/OFF): Either by a software Menu or by an external signal, applied via Digital Input 7, if such is installed (see INSTALLED OPTIONS menu). Emerson Process Management GmbH & Co. OHG...
Page 172
The next line "Signal Range" allows to configure the analog outputs ranges / the analog outputs to comply with the NAMUR NE43 recom- mendations as described on the next page. Emerson Process Management GmbH & Co. OHG...
Page 173
Table 6-1 at page 6-9 shows a comparison of NE43 gives recommendations how to setup all available modes. analog outputs to avoid above situation and the X-STREAM series analyzers consider NE 43: Emerson Process Management GmbH & Co. OHG...
Dual channel instrument: Pressing the key while any line is selected returns to the optional gas component selec- tion menu to open the same menu for the other measuring channel. 6-10 Emerson Process Management GmbH & Co. OHG...
Page 175
PressSensor:Internal Wrong entries may cause faulty Flow monitor: measuring results or strange analyzer behaviour! When entering this menu for the first time just use it to get informed about the instrument's configuration. Emerson Process Management GmbH & Co. OHG 6-11...
The moment the measured concentration Level2: 500 ppm exceeds one of the thresholds, a status message Function: High appears in the measuring screen's 4 line and the related digital output (option) is set. 6-12 Emerson Process Management GmbH & Co. OHG...
0 ppm 1000 ppm Example 1 Example 2 110% 100 ppm -100 ppm +1100 ppm Table 6-2: Thresholds influenced 220% 120% 1200 ppm -1200 ppm +2200 ppm Example 3 by SpanRange parameter Emerson Process Management GmbH & Co. OHG 6-13...
Page 178
Alarm signal level is digital "LOW". Note! "Function" parameter factory setting is Off FS (fail-safe) unless otherwise specified at time of order. 6-14 Emerson Process Management GmbH & Co. OHG...
Level 1 function: High FS • Level 2 function: Low FS As long as any alarm is activated a status message appears in the measuring screen's line. Fig. 6-1: Thresholds defining a window Emerson Process Management GmbH & Co. OHG 6-15...
Level 1 function: High FS Level 2 function: High FS • As long as any alarm is activated a status message appears in the measuring screen's line. Fig. 6-2: HIGH and HIGH-HIGH alarm mode 6-16 Emerson Process Management GmbH & Co. OHG...
Level 1 function: Low FS • Level 2 function: Low FS • As long as any alarm is activated a status message appears in the measuring screen's line. Fig. 6-3: LOW and LOW-LOW alarm mode Emerson Process Management GmbH & Co. OHG 6-17...
Page 182
This behaviour may cause trouble when the instrument is connected to e.g. a data aquisition system. 6-18 Emerson Process Management GmbH & Co. OHG...
Page 183
Save-Load-Update CfgData>COMport! COMport>CfgData.. Verify! menu page Press the DOWN key to open the second menu page. Now select the "CfgData>UserData.." line and FactData>CfgData.. press the ENTER key. CfgData>UserData.. UserData>CfgData.. menu page Emerson Process Management GmbH & Co. OHG 6-19...
Page 184
COMMAND EXECUTED screen shows up. Note! For a detailed description of all SAVE-LOAD- UPDATE options 7-6, page 7-34. Press the HOME key to return to the MEASURING SCREEN, because the setup checking procedure is finished. 6-20 Emerson Process Management GmbH & Co. OHG...
Dual channel instruments: 5 minutes! Analyzers with parallel tubing require separate leak tests for each gas path ! Emerson Process Management GmbH & Co. OHG...
Advanced calibration is a more comfortable 7-3-3, pg. 7-10), single automatic calibrations variation of manual calibration, providing ONE 7-3-4, pg. 7-20) and unattended KEY calibrations supported by internal and/or automatic calibrations ( 7-3-4-4, pg. 7-28). external valves. Emerson Process Management GmbH & Co. OHG...
When done, press the LEFT key to return to the CALIBRATION menu. Emerson Process Management GmbH & Co. OHG...
Page 189
Emerson Process Management GmbH & Co. OHG...
Page 190
Tol.Check: "none" (see installed options menu) and are Purge time: 15 s used for auto calibrations and unattended Interval time... calibrations only (see related sections for a description). Emerson Process Management GmbH & Co. OHG...
Select the second line to start the calibration. START calibration! Line 3 shows the calibration gas setup (here: ZeroGas 0.000 ppm required zero gas concentration is 0.000 ppm), CO2.1 0.200 ppm while line 4 shows the currently measured gas concentration. Emerson Process Management GmbH & Co. OHG...
Select the second line to start the calibration. START calibration! Line 3 shows the calibration gas setup (here: SpanGas 20.000 ppm required span gas concentration is 20 ppm), CO2.1 16.200 ppm while line 4 shows the currently measured gas concentration. Emerson Process Management GmbH & Co. OHG...
Page 193
HOME key to return to the measure- ment screen to close the manual calibration procedures. Emerson Process Management GmbH & Co. OHG...
1 ch 2 ch 2 ch 2 ch 1 ch 1 ch 1 improvement in & & & Zero spanning with time variable assignment timeline Fig. 7-2: Calibration improvement by variable valve assignments Emerson Process Management GmbH & Co. OHG 7-11...
Page 196
V4 V2 V4 V1 V2 V1 Dual channel analyzer, both channels • require the same zero AND span gas concentrations: Possible assignments Span V1 V2 V1 V4 V2 V4 Zero V2 V1 V4 V1 V4 V2 7-12 Emerson Process Management GmbH & Co. OHG...
Page 197
Depending on the gases used this may allow selecting the same combination for both • higher calibration performance . channels selecting combinations where one valve • has the same function for both channels Emerson Process Management GmbH & Co. OHG 7-13...
"ALL" relates to the single channel only! The analyzer immediately begins zero calibration(s). Watch the screen for status information (explained by means of exemplary pictures on the next page). 7-14 Emerson Process Management GmbH & Co. OHG...
Page 199
0 seconds and the gas flow is sample. Now press either the LEFT key to return to the advanced calibration menu to select another calibration procedure or the HOME key to return to the measuring screen. Emerson Process Management GmbH & Co. OHG 7-15...
"ALL" relates to the single channel only! The analyzer immediately begins span calibration(s). Watch the screen for status information (explained by means of exemplary pictures on the next page). 7-16 Emerson Process Management GmbH & Co. OHG...
Page 201
(see page 7-6). The SpanAll calibration procedure has finished when the last time interval shows remaining 0 seconds and the gas flow is sample. Press the HOME key to return to the measuring screen. Emerson Process Management GmbH & Co. OHG 7-17...
The analyzer immediately begins zero calibration(s for all channels), followed by span calibration(s). Watch the screen for status information (explained by means of exemplary pictures on the next page). 7-18 Emerson Process Management GmbH & Co. OHG...
Page 203
Procedure Purging with the related gas only. Postpurge time again Time 10 s is the purge time entered in the calibration setup screen (see page 7-6). Emerson Process Management GmbH & Co. OHG 7-19...
Page 204
(see page 7-6). The SpanAll calibration procedure has finished when the last time interval shows remaining 0 seconds and the gas flow is sample. Press the HOME key to return to the measuring screen. 7-20 Emerson Process Management GmbH & Co. OHG...
1 ch 2 ch 2 ch 2 ch 1 ch 1 ch 1 improvement in & & & Zero spanning with time variable assignment timeline Fig. 7-3: Calibration improvement by variable valve assignments Emerson Process Management GmbH & Co. OHG 7-21...
Page 206
V4 V2 V4 V1 V2 V1 Dual channel analyzer, both channels • require the same zero AND span gas concentrations: Possible assignments Span V1 V2 V1 V4 V2 V4 Zero V2 V1 V4 V1 V4 V2 7-22 Emerson Process Management GmbH & Co. OHG...
Page 207
Depending on the gases used this may allow selecting the same combination for both • higher calibration performance . channels selecting combinations where one valve • has the same function for both channels Emerson Process Management GmbH & Co. OHG 7-23...
Select the second line to start the calibration. ZeroGas 0.000 ppm Line 3 shows the calibration gas setup (here: CO2.1 0.200 ppm required zero gas concentration is 0.000 ppm), while line 4 shows the currently measured gas concentration. 7-24 Emerson Process Management GmbH & Co. OHG...
Page 209
(see page 7-6). The zero calibration procedure has finished when the last time interval shows remaining 0 seconds and the gas flow is sample. Emerson Process Management GmbH & Co. OHG 7-25...
20.000 ppm), while line 4 shows the currently measured gas concentration. After having started a span calibration watch the screen for status information (explained by means of the left hand exemplary pictures): 7-26 Emerson Process Management GmbH & Co. OHG...
Page 211
(dual channel analyzer only), if required select the second channel and perform the steps above to span calibrate this channel, too, press the HOME key to return to the measure- ment screen. Emerson Process Management GmbH & Co. OHG 7-27...
Ensure the warm-up time after is done for both channels! switching on has elapsed! Warm-up time is 15 to 50 minutes depending on installed measu- ring system and configuration! 7-28 Emerson Process Management GmbH & Co. OHG...
DOWN key to open the MAIN MENU, enter the CONTROL.. menu. Zero calibration.. Enter the last line (APPLY GAS..) Span calibration.. Apply gas.. Dual channel instrument: Optional gas component selection menu - Select the component to be setup 7-30 Emerson Process Management GmbH & Co. OHG...
Cancelling during postpurge does not influ- ence the procedure because the new data has already been calculated and stored, and the (post-)purge time cannot be shortened (except by changing the related setup menu parameter). Emerson Process Management GmbH & Co. OHG 7-31...
Page 216
Sample bration was cancelled: As the sample valve CO2.1 0.000 ppm was opened again, a postpurge procedure was Procedure Purging started. Time 10 s Press the LEFT key to exit these screens. 7-32 Emerson Process Management GmbH & Co. OHG...
CrossInterference (Pure span gases are only required for this special calibration procedure. Further Status: calibrations can be performed with standard @ Span Cal: span gases). Emerson Process Management GmbH & Co. OHG 7-33...
Page 218
STATUS parameter is Pure span gases are not longer required and for cost saving reasons may now be replaced by standard span gases. 7-34 Emerson Process Management GmbH & Co. OHG...
Temporary flushing with Nitrogen in Oxygen concentration. (Operation at 40 °C ) for less than 1 h (e.g. for analyzer zeroing halves lifetime). purpose) has no influence on measuring characteristics. Emerson Process Management GmbH & Co. OHG 7-35...
Page 220
(e.g. instruments with earth connectors, heating installations). This should be done periodically when working at open instruments (especially after leaving the service site, because e.g. walking on low conducting floors might cause additional ESD). 7-36 Emerson Process Management GmbH & Co. OHG...
Do not disassemble or repair the sensor. Removing a sensor part or remodeling the sensor will damage the sensor or leak the electrolyte and restoration to the original condition may not be possible. Emerson Process Management GmbH & Co. OHG 7-37...
Page 222
Swallow a large amount of tap water or 600 cc of milk and vomit it. Consult a doctor promptly. Discarded sensors may cause environmental pollution. Return a worn-out sensor to Emerson Process Management or an industrial waste management contractor when discarding a worn-out sensor. 7-38...
Flip down the front door carefully to not damage the instrument, hinges or equipment installed below the analyzer. Fig. 7-4: X-STREAM interior views (covers and doors not shown) Emerson Process Management GmbH & Co. OHG 7-39...
Detent springs Fig. 7-6: Cardcage Detail Press here to open Detent springs sensor unit Detent springs Fig. 7-8: Purged Box Detail Fig. 7-7: Allocation of eO Sensor Unit (BKS removed) 7-40 Emerson Process Management GmbH & Co. OHG...
Take the new sensor, remove it's plug, place it into the sensor block and attach the connector to the related strip P2 on the electronics board. Sensors Fig. 7-10: Sensor Block Assembly Emerson Process Management GmbH & Co. OHG 7-41...
Once the output signal has been adjusted for Potentiometer R4 a specific sensor changing the potentiometer Strip P2 for Oxygen sensor signal settings will cause incorrect measuring results! Fig. 7-11: OXS Board, top view 7-42 Emerson Process Management GmbH & Co. OHG...
Send it back to the EMERSON Process Management factory (or to your local sales office) or to an industrial waste management contractor for waste disposal. Emerson Process Management GmbH & Co. OHG 7-43...
PVC caps to avoid pollution of inner gas path. • Moisten the lint-free cloth with a mixture of 3 parts of water and 1 part of the general purpose detergent. 7-44 Emerson Process Management GmbH & Co. OHG...
For information about how to: save CfgData to UserData 7-6-1, page 7-35 restore UserData to CfgData 7-6-2, page 7-36 restore FactData to CfgData 7-6-3, page 7-37 save / restore CfgData to/from COMPort 7-6-4, page 7-38 Emerson Process Management GmbH & Co. OHG 7-45...
Any further changes affecting the instrument's setup update the CfgData only, as long as not again stored into the UserData set. When the procedure has finished, the -COMMAND EXECUTED- COMMAND EXECUTED screen shows up. 7-46 Emerson Process Management GmbH & Co. OHG...
- PLEASE WAIT - parate memory area. The restored data is then called CfgData and used for operating the Procedure X:078000 analyzer. When the procedure has finished, the COMMAND EXECUTED screen shows up. -COMMAND EXECUTED- Emerson Process Management GmbH & Co. OHG 7-47...
- PLEASE WAIT - memory area. The restored data is then called CfgData and used for operating the analyzer. Procedure X:078000 When the procedure has finished, the COMMAND EXECUTED screen shows up. -COMMAND EXECUTED- 7-48 Emerson Process Management GmbH & Co. OHG...
Changing this parameter causes trouble if not supported by the hardware! Serial interface baud rate. Baud rate: 19200 Available options: Parity: None 2400, 4800, 9600, 19200 Supported parity bit. Available options: menu page None, Even, Odd Emerson Process Management GmbH & Co. OHG 7-49...
- PLEASE WAIT - (and possibly changed by operator) analyzer setup on an external device for future use. Procedure X:078000 When the procedure has finished, the COMMAND EXECUTED screen shows up. -COMMAND EXECUTED- 7-50 Emerson Process Management GmbH & Co. OHG...
Page 235
The analyzer now loads the data from the Verify! external device and compares it to the CfgData. A status screen comes up followed by a COMMAND EXECUTED screen or an error message, if the data does not match. Emerson Process Management GmbH & Co. OHG 7-51...
Procedure X:078000 The restored data is then called CfgData and used for operating the analyzer. When the procedure has finished, the COMMAND EXECUTED screen shows up. -COMMAND EXECUTED- 7-52 Emerson Process Management GmbH & Co. OHG...
(e.g. frozen) due to the ongoing work on the instrument (e.g. during calibration). If solving a reported problem requires working inside an open instrument take care of the safety instructions given at the beginning of this manual! Emerson Process Management GmbH & Co. OHG...
Page 239
4. After calibration the sample time and 2 times t to elapse. gas valve has opened, but the If appropriate reduce purge time interval of purge time and time. 2 times t has not yet elapsed Emerson Process Management GmbH & Co. OHG...
Page 240
2. Wait for instrument to heat up 2. Temperature of analyzer requires tempe- components or physical rature monitoring check internal heater for thermostatted compartment is enabled within the proper function not within the configured range service level Emerson Process Management GmbH & Co. OHG...
Page 241
Explanation: Concentration alarm level 1 is activated (exceeded) Alarm Level2 Alarm level 2 was exceeded Adjust the gas concentration to be within the set limits Explanation: Concentration alarm level 2 is activated (exceeded) Emerson Process Management GmbH & Co. OHG...
Page 242
10 % from setpoint) (first calibration after restarting the calibration installation) Spanning Span calibration ongoing for the Wait until calibration has finished Check function channel identified by the tag Cancel calibration Explanation: Ongoing span calibration Emerson Process Management GmbH & Co. OHG...
Page 243
10 % from setpoint) (first calibration after restarting the calibration installation) Zeroing Zero calibration ongoing for the Wait until calibration has finished Check function channel identified by the tag Cancel calibration Explanation: Ongoing zero calibration Emerson Process Management GmbH & Co. OHG...
Nuts to loosen (two on both sides of the front panel) Fig. 8-0: X-STREAM F interior view with flapped front panel Emerson Process Management GmbH & Co. OHG...
Page 245
Remove restriction behind gas outlet Reduce gas flow or pump rate 4. Sensor or detector not 4. Check detectors connections connected 5. Electrochemical Oxygen 5. Check sensor and replace if sensor worn-out need be Emerson Process Management GmbH & Co. OHG...
Page 246
Sensor failure ( status message "PressSensor“, page 8-4) 10. Check temperature of gas path(s) Remove all sources of condensation Keep all temperatures at least 10 °C above sample gas temperature 11. Replace BKS 8-10 Emerson Process Management GmbH & Co. OHG...
Page 247
Replace solenoid valves For valve control via serial interface or digital inputs: Any valve activated? 5. Check gas path and sample 5. Gas path(s) polluted handling system for pollution Clean gas path Emerson Process Management GmbH & Co. OHG 8-11...
Flip down the front door carefully to not damage the instrument, hinges or equipment installed below the analyzer. Fig. 8-1: X-STREAM interior views (covers and doors removed) 8-12 Emerson Process Management GmbH & Co. OHG...
No or wrong signal How to solve: Potential cause: Check supply voltage +6V voltage faulty Replace BKS board +3 V Reference voltage faulty (measure at C179, pin 1 (lower pin in fig. 8-2)) Emerson Process Management GmbH & Co. OHG 8-13...
Oxygen measurement w/o IR channel: Signal level set point: LOW < 0.45 V; HIGH > 2.4 V (TTL logic) µP not working Switch off analyzer and on again BKS board defective Replace BKS board Emerson Process Management GmbH & Co. OHG 8-15...
Replace BKS board Note! Sensor is worn-out when, connected to ambient air, the output voltage is less than 2.8 V: Replace sensor! Tp 2 Tp 1 Fig. 8-4: OXS board, assembled, top view Emerson Process Management GmbH & Co. OHG 8-17...
If J7 is not set properly all data gets lost and the analyzer is reset to factory default settings the moment the analyzer is disconnected from power supply! Battery buffer jumper Fig. 8-5: BKS board (section) 8-18 Emerson Process Management GmbH & Co. OHG...
Use only the same type and ratings of fuse! Fuse data: Type Wickmann 372 T 4 A / 250 V Fig. 8-6: Allocation of fuse on BKS board Emerson Process Management GmbH & Co. OHG 8-19...
Modbus_over_serial_line_V1.pdf. For a list of supported functions 9-2, page 9-2 supported parameters and registers, ordered by parameter name 9-2, page 9-3 comparison of registers and parameters, ordered by register number 9-3, page 9-16 Emerson Process Management GmbH & Co. OHG...
Daniel/Enron Modicon Data type 8001 - 8499 equal 5001 - 5999 long word 9001 - 9999 equal 6001 - 7999 floating point the following pages for comparisons of all Daniel/Enron and Modicon registers. Emerson Process Management GmbH & Co. OHG...
Dyn, Cfg if internal pressure measurement is enabled 9025 6049..6050 500 .. 2000 (PressureSensorInstalledDB) this is a dynamic Float variable; if no pressure sensor is installed we have to input the current value Emerson Process Management GmbH & Co. OHG...
Page 260
0 = access is allowed 3115 3115 1 = access requires user code mode for getting access to basic areas 2 = access is prohibited enum BoardSerialNrDB Cfg, Fac 3101 3101 computer board serial number Word Emerson Process Management GmbH & Co. OHG...
Page 261
CrossInterferenceFact#DB Cfg, Self 9089 6177..6178 9090 6179..6180 factor that determines the cross interference Float influence between 2 measurement channels CrossInterferenceOffset#DB Dyn, Sim 9009 6017..6018 9010 6019..6020 zero correction of cross interference Float compensation Emerson Process Management GmbH & Co. OHG...
Page 262
2015 = 1/0 2016 = 1/0 2016 = 1/0 switch on/off span gas valve of channel dbSwitchValves (VALVE_ZEROGAS) 2013 = 1/0 2013 = 1/0 2014 = 1/0 2014 = 1/0 switch on/off zero gas valve Emerson Process Management GmbH & Co. OHG...
Page 263
(0 = Off) Integer 2 characters per register: EmersonAccessCodeDB Cfg, Fac 3126..3129 3126..3129 character at HIGH byte, user code for getting access to Emerson areas String-8 character at LOW byte. EmersonAccModeDB 0 = access is allowed Cfg, Fac 3130 3130...
Page 264
Boolean Limit2Level#DB Limit2Level#DB unit is ppm; 9102 6203..6204 min/max depend on EndOfRange#DB, concentration is compared with this level; 9104 6207..6208 MaxConcePercent#DB, according Limit2AlarmTyp#DB an alarm Float DifferentialMeasurementDB contact gets activated Emerson Process Management GmbH & Co. OHG...
Page 265
MaxValue#DB Cfg, Self 9019 6037-6038 9020 6039..6040 maximum offset corrected raw value at end of Float range MeasBufAvg#DB Dyn, Sim 9099 6197..6198 9100 6199..6200 average value of used ring buffer values Float Emerson Process Management GmbH & Co. OHG...
Page 266
6125..6132 polynom coefficients that compute span factor A0, A1, A2, A3 9067..9070 6133..6140 correction (PressureFactor#DB) using the Float-Arr4 assigned pressure (PressureSensorInstalledDB) PresDecimalPointDB 3090 3090 0..4 decimal point position for pressure displays uchar 9-10 Emerson Process Management GmbH & Co. OHG...
Page 267
1 = digital input controls pump Enum instance PumpInstalledDB Cfg, Fac 0 = not installed 3017 3017 1 = installed device has a pump installed and controls it Boolean Emerson Process Management GmbH & Co. OHG 9-11...
Page 268
Boolean remote access' RemoteSecurityDB 3008 3008 sets ParamAccessModeDB to 'service access' Word if input value equals service code ResponseTime#DB 3002 3002 2..60 3003 3003 signal damping (in secs) for gas change Integer 9-12 Emerson Process Management GmbH & Co. OHG...
Page 269
Enum StartOfRange#DB 9031 6061..6062 9032 6063..6064 beginning of measurement range in ppm Float TcntCorrCoeff#DB Cfg, Fac 9085 6169..6170 9086 6171..6172 gradient m of equation 'RawValueTfact#DB = Float m*Temperature + 1'; see RawValueTfact#DB Emerson Process Management GmbH & Co. OHG 9-13...
Page 270
(TemperatureFactor#DB) using the Float-Arr4 assigned temperature (TempCompSpan#DB) ToffCorrCoeffs#DB Cfg, Fac 9047..9050 6093..6100 polynom coefficients that compute zero offset A0, A1, A2, A3 9051..9054 6101..6108 correction (TemperatureOffset#DB) using the Float-Arr4 assigned temperature (TempCompZero#DB) 9-14 Emerson Process Management GmbH & Co. OHG...
Page 271
Fac: parameter, only setable by service/factory people Self: parameter, determined by device's algorithms itself Sim: can be simulated, then writable Fct: function to be called Ver: parameter, set by version control Rd: only readable Wr: only writable Emerson Process Management GmbH & Co. OHG 9-15...
Enclose a cover letter and purchase or- returned as directed by the purchase order and der and ship the defective equipment ac- shipping instructions. cording to instructions provided in a Rosemount Return Authorization, prepaid, Emerson Process Management GmbH & Co. OHG 10-1...
In Asia Pacific: In Asia Pacific: Emerson Process Management Emerson Process Management Asia Pacific Pte Limited Asia Pacific Pte Limited 1 Pandan Crescent 1 Pandan Crescent Singapore 128461 Singapore 128461 +65-6-777-8211 +65-6-777-8211 10-2 Emerson Process Management GmbH & Co. OHG...