Status Setting List; When Writing Status - Ulvac G-TRAN Series Instruction Manual

Multi ionization gauge sensor unit
Hide thumbs Also See for G-TRAN Series:
Table of Contents

Advertisement

15.4.

Status setting list

This shows the setting of the filament on and the degas on, errors, and setpoint status.

When writing status

Filament ON/OFF signal
*Note*
The use of the SH200 filament ON/OFF signal differs between the
SH200 stand-alone mode and the combination mode.
SH (Status/high bit)
bit
Status
B7
FIL 1/2
B6
FIL ON/OFF
B5
N・C
B4
DEGAS ON/OFF
FIL 1/2
※SH200 stand-alone mode FIL ON/OFF
※Combination mode FIL OFF/ON
DEGAS ON/OFF
SL (Status/low bit)
bit
Status
B3
N・C
B2
N・C
B1
N・C
B0
N・C
※ Both [1] and [0] work for N・C.
Example) When turning on the filament using the filament1, the status is as
follows:
bit
B7
B6
B5
B4
Therefore, when the address "11", the command to be sent is as follows:
command
:
:
0
8
〔30H〕
〔38H〕
0
1
0
0
0
0
0
0
0〔30H〕
0
0
0
0
Status
C〔43H〕
FIL 1
1
FIL ON
1
N.C
0
DEGAS ON
0
AD0
AD1
SW
1
1
SW
C
4
〔43H〕
〔34H〕
1
0
1
1
0
0
0
0
1:FIL1
0:FIL2
1:ON
0:OFF
1:OFF
0:ON
1:ON
0:OFF
bit
Status
B3
N.C
B2
N.C
B1
N.C
B0
N.C
SH
SL
CHKH
C
0
44
D
5
〔44H〕
〔35H〕
1
0
1
1
0
0
1
1
0〔30H〕
0
0
0
0
CHKL
CR
7
7
CR