Sh200 Stand - Alone Mode Standard; Specifications; Standard Specifications - Ulvac G-TRAN Series Instruction Manual

Multi ionization gauge sensor unit
Hide thumbs Also See for G-TRAN Series:
Table of Contents

Advertisement

2.3.
SH200 stand-alone mode standard specifications
Measurable
pressure range
(N
)
2
Accuracy (N
)
2
POWER LED status
Control input
signal
2.4.
SWU combination mode (SH200+SWU) standard specifications
Measurable
pressure range
(N
)
2
Accuracy (N
)
2
POWER LED status
Control input
signal
-8
5.0×10
Pa to 1.0×10
-8
5.0×10
Pa to 1.0×10
White
: Start up
Blue
: Normal operation
Green
: Filament is on
Blinking Red: Power failure
FIL ON/OFF, FIL 1/2
DEGAS ON/OFF
Open collector input,
Negative logic
-8
5.0×10
Pa to 1.0×10
When pressure drops: SWU indicates 2Pa or lower, SWU ⇒SH200
When pressure rises: SWU indicates 3Pa or higher, SH200⇒SWU
※ SH200 measurement can be forcibly turned off with the
control signal.
Pressure range
-8
5.0×10
Pa to 4.0×10
-1
4.0×10
Pa to 3.0×10
+0
3.0×10
Pa to 1.0×10
+4
1.0×10
Pa to 1.0×10
In the overlapping pressure range of 0.4Pa to 3.0Pa, the
measured pressure of the Pirani gauge and SH200 is adjusted and
output.
※ SWU and SH200 have gas species dependency. Please note
the difference in pressure indication when gauges switch.
White
: Start up
Blue
: Normal operation to SWU measurement range
Green
: SH200 filament is on
Red
: SWU power failure
Blinking Red: SH200 power supply voltage error
FIL ON/OFF, FIL 1/2, DEGAS ON/OFF
Open collector input, Negative logic
※ When FIL ON/OFF signal is Lo input, the filament of SH200
is off.
+1
Pa
+1
Pa: ±15%
+5
Pa
Accuracy
-1
Pa
±15%
+0
Pa
±10% to ±15%
+4
Pa
±10%
+5
Pa
±20%
6