4.4.
I/O connector serial communication type SH200-R
1
9
Fig. 4-4 Serial communication type SH200-R I/O connector pin layout
Pin
This unit
number
1
Power supply
2
3
4
RS232C RxD
Terminating resistor
5
for RS485
6
RS232C TxD
7
8
Measured value output
9
Power supply GND
10
RS485-
11
12
RS485+
RS485+
13
(For connecting the
terminating resistor)
14
RS232C GND
15
Signal GND
Case
Frame ground
Wiring
*Note*
The shaded pins are used in the internal circuit, so do not wire them.
(D-sub15 pin(male), M2.6 screw)
Power supply for driving this unit
RxD of RS-232C
Terminating resistor for RS-485.
It is connected to pin13.
TxD of RS-232C
It outputs a pressure signal.
Ground for the power supply
Minus of RS-485
Plus of RS-485
Terminating resistor for RS-485.
It is connected to pin5.
Ground for RS-232C
Ground for the output signal
Ground for frame
15
8
15
Function
(DC20V to 28V)
(DC0V to 10V)