Adjustment Command; And Zero Point Adjustment Initialization Command; Checking The Value Of Filament Power Monitoring Of Sh200; Reading Setpoint1 Value - Ulvac G-TRAN Series Instruction Manual

Multi ionization gauge sensor unit
Hide thumbs Also See for G-TRAN Series:
Table of Contents

Advertisement

SAU/SWU/SPU zero point adjustment command
Command
:
AD0
SAU zero point adjustment condition
SWU/SPU zero point adjustment condition: SWU/SPU pressure is about 2Pa or
When the zero point is adjusted, 'o' is returned.
When the pressure is beyond the range of the zero point adjustment, 'n' is
returned.
SAU/SWU atmospheric pressure and zero point adjustment initialization command
Command
:
AD0
When initialization is performed, 'o' is returned.
When the initialization fails, or no compatible device is connected, 'n' is
returned.

Checking the value of filament power monitoring of SH200

Command
:
AD0
Return format from this unit to PC.
: AD0
AD1
FIL
The value that is set the maximum power supplied to the SH 200 filament as
100% is returned to "XXX".
This value is larger when the emission current is at 1mA than at 10μA, and
when the pressure is higher. The value also varies depending on the
measuring gas.
The maximum value may not be 100% due to variations in the power supply of
the SH200.
If the maximum value is more than 90% or less than 20%, it is nearing the
end of the filament.
Reading Software Version
Command
:
AD0
Return format from this unit to PC.
: AD0
AD1
T
"SH200A" indicates the model name. "315" indicates the software version,
Ver. 3.15.
The software version is subject to change without notice.

Reading Setpoint1 Value

Command
:
AD0
Return format from this unit to PC.
: AD0
AD1
1
"X.XXE±XX" contains the setpoint value. "±" can be "+" or "-".
AD1
ZER
CHKH
AD1
CLR
CHKH
AD1
FIL
CHKH
X
X
X
CHKH
AD1
T
CHKH
S
H
2
0
0
AD1
1R
CHKH
X
.
X
X
E
40
CHKL
CR
: SWU/SPU pressure is 1,000Pa or
lower.
lower.
-3
(1.0×10
Pa or lower recommended)
CHKL
CR
CHKL
CR
CHKL
CR
CHKL
CR
A
3
1
5
CHKH
CHKL
CR
±
X
X
CHKH
CHKL
CR
CHKL
CR