Baud Rate Setting; Address Setting - Ulvac G-TRAN Series Instruction Manual

Multi ionization gauge sensor unit
Hide thumbs Also See for G-TRAN Series:
Table of Contents

Advertisement

RS-485 with terminating resistor (Example)
Example) USB serial interface USB-485 manufactured by NATIONAL INSTRUMENTS.
If the number of RS-485 connections is large, the total length of the connection
cable is more than 15m, or communication errors frequently occur, install a
terminating resistor in the terminating equipment.
This unit Pin No.
Terminating
resistor

Baud rate setting

The baud rate can be set on the G-TRAN setting tools "UL-MOBI_Android" or "UL-
MOBI_Windows" by connecting a PC or smartphone to the "USB connector" of the main unit.
Please refer to the application manual for details.
The configurable baud rates are "9600bps", "19200bps", and "38400bps".

Address setting

The Address can be set on the G-TRAN setting tools "UL-MOBI_Android" or "UL-
MOBI_Windows" by connecting a PC or smartphone to the "USB connector" of the main unit.
Please refer to the application manual for details.
The setting range is from 01 to 31.
The maximum number of connections per one line for RS-485 is 31 including
the host.
The setting is valid at any time.
This address number is also used for RS-232C.
Download:
UL-MOBI_Android Google Play
UL-MOBI_Windows
UL-MOBI_Windows Instruction manual
RS485+
12
RS485-
10
GND
14
5
RS485+
13
https://www.ulvac.co.jp/download/en/application/
https://www.ulvac.co.jp/download/en/application/
Host side
4
8
5
9
1
37
RxD+
TxD+
RxD-
TxD-
GND