Setting Of Auto Calibration Component/Range - Fuji Electric ZPAJ Instruction Manual

Ndir type infrared gas analyzar
Table of Contents

Advertisement

6.2.4 Setting of auto calibration component/range

ENT
performed.
"Auto Calibration Component/range"
setting
Auto calibration and the manual calibration
of the component for which "AR" has been
selected as range switch mode are performed
in the range selected here. In this case, once
the calibration is started, the range is
automatically switched, and on completion
of the calibration, the original range is
resumed.
The range identification contact is inter-
locked with the range after the switch.
However, if the hold setting is set to "ON,"
the contact status before calibration is main-
tained.
INZ-TN2ZPAJ-E
-
-
key.
ENT
key.
ENT
key.
33
Cal. Settings
Auto Cal.
-
Ch1
Range1
NOx
Range2
Ch2
Range1
SO
Range2
2
Ch3
Range1
CO
Range2
Ch4
Range1
O
Range2
2
End of Auto Calibration
Cal. Settings
Auto Cal.
Ch1
Range1
Range2
NOx
Ch2
Range1
SO
Range2
2
Ch3
Range1
CO
Range2
Range1
Ch4
O
Range2
2
End of Auto Calibration
component setting
Select a range for
auto calibration
0–500.0
ppm
enable
0–5000
ppm
0–500.0
ppm
enable
0–2500
ppm
0–200.0
ppm
enable
0–2000
ppm
0–10.00
vol%
disable
0–25.00
vol%
(
)
ENT
Range Setting
Set enable or disable
for auto calibration
0–500.0
ppm
enable
0–5000
ppm
0–500.0
ppm
enable
0–2500
ppm
0–200.0
ppm
enable
0–2000
ppm
0–10.00
vol%
disable
0–25.00
vol%
(
)
ENT
-