6.2.4 Setting of auto calibration component/range

Select the Ch (component) and the range for which auto calibration is to be performed. The Ch for
which "AR" has been selected as range switch mode is calibrated in the range set here. Auto cali-
bration and the manual calibration of the component for which "AR" has been selected as range
switch mode are performed in the range selected here.
(1) Select < User mode >  < Calibration
parameters >  < Auto calibration com-
ponent/range >. "Auto Calibration Com-
ponent Range" setting screen appears as
shown at right.
(2) Select the Ch you want to change by
pressing the
ENT
key and the selected cursor is high-
lighted.
(3) Select the range to be calibrated mainly
by pressing the
(4) Then press the
is performed in the selected range when
auto calibration or auto zero calibration is
performed.
"Auto Calibration Component/range"
setting
Auto calibration and the manual calibration
of the component for which "AR" has been
selected as range switch mode are performed
in the range selected here. In this case, once
the calibration is started, the range is
automatically switched, and on completion
of the calibration, the original range is
resumed.
The range identification contact is inter-
locked with the range after the switch.
However, if the hold setting is set to "ON,"
the contact status before calibration is main-
tained.
(5) Press the
(3), and the highlight is switched between
"enable" and "disable" auto calibration.
(6) Select "enable" or "disable" by pressing
the
or the
(7) Then press the
INZ-TN2ZPA-E
or the
key. Press the
or the
key.
ENT
key, and calibration
key in the state described in
key.
ENT
key.
End of Auto Calibration
End of Auto Calibration
component setting
33
ENT
(
)
Range Setting
ENT
(
)