6.2.4 Setting of auto calibration component/range

ENT
performed.
"Auto Calibration Component/range"
setting
Auto calibration and the manual calibration
of the component for which "AR" has
been selected as range switch mode are
performed in the range selected here. In this
case, once the calibration is started, the range
is automatically switched, and on completion
of the calibration, the original range is
resumed.
The range identification contact is inter-
locked with the range after the switch.
However, if the hold setting is set to "ON,"
the contact status before calibration is main-
tained.
key.
ENT
key.
ENT
key.
28
Cell. Settings
Select a range for
Auto Cal.
auto calibration
Ch1
Range1
0–20.00
CH
Range2
0–100.0
4
Range1
0–20.00
Ch2
CO
Range2
0–100.0
2
Ch3
Range1
0–500.0
H
S
Range2
0–2000
2
Ch4
Range1
0–10.00
O
Range2
0–25.00
2
(
End of Auto Calibration
Range Setting
Cell. Settings
Set enable or disable
Auto Cal.
for auto calibration
Ch1
Range1
0–20.00
CH
Range2
0–100.0
4
Range1
0–20.00
Ch2
CO
Range2
0–100.0
2
Ch3
Range1
0–500.0
H
S
Range2
0–2000
2
Ch4
Range1
0–10.00
O
Range2
0–25.00
2
(
End of Auto Calibration
component setting
vol%
enable
vol%
vol%
enable
vol%
ppm
enable
ppm
vol%
enable
vol%
ENT
)
vol%
enable
vol%
vol%
enable
vol%
ppm
enable
ppm
vol%
enable
vol%
ENT
)