4.10.6. How to compensate the measurement value
Description
Measurement value can be calibrated arbitrarily.
Zero point and span adjustment can be made.
Settable range
(1) Zero point : -5 to +5 [m/s] in terms of flow velocity in
piping.
(2) Span
: ±200%
The output value (reading, analog output and total output)
is computed by the following expression.
Measurement value × [Span set value %]
Output =
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.)
Operation
Compensate the zero point to 0.5m
(example)
Key operation
Press the
▼
Press the
ENT
▼
Press the
▼
Press the
ENT
▼
Press the
▼
Press the
▼
Press the
ENT
▼
▼
――― "CALIBRATION ZERO" has been registered. ―――
▼
▼
Press the
▼
Press the
ENT
▼
Press the
▼
Press the
▼
Press the
ENT
▼
▼
――― "CALIBRATION SPAN" has been registered. ―――
▼
▼
ESC
Press the
the measurement mode.
INF-TN3FLR-E
+ Zero point
100
3
/h, and the span by +1%.
Description
key twice to display "OUTPUT SETUP".
key once to display "ZERO ADJUSTMENT".
ENT
key for 8 times to display "CALIBRATION ZERO".
key once to blink the cursor.
ENT
key for 6 times to move the cursor.
key for 5 times to set "5".
key once to register.
ENT
key once to display "CALIBRATION SPAN".
key once to blink the cursor.
ENT
key twice to move the cursor.
key once to set "1".
key once to register.
ENT
ESC
key once and then press the
Output
Flow rate
0
Zero adjustment movement
key for 3 times to enter
Output
100%
Flow rate
0
Span movement
Display
OUTPUT SETUP
ZERO ADJUSTMENT
SET ZERO
CALIBRATION ZERO
0.000 m3/h
CALIBRATION ZERO
00000.000 m3/h
CALIBRATION ZERO
00000.000 m3/h
CALIBRATION ZERO
00000.500 m3/h
CALIBRATION ZERO
COMPLETE
↓
CALIBRATION ZERO
0.500 m3/h
CALIBRATION SPAN
100.0 %
CALIBRATION SPAN
100.0 %
CALIBRATION SPAN
100.0 %
CALIBRATION SPAN
101.0 %
CALIBRATION SPAN
COMPLETE
↓
CALIBRATION SPAN
101.0 %
0.000
%
0.000
m3
- 59 -