Pc Block; General Procedures Of Bep Assy Disassembly; Bep (Back End Processor) Assy(Fru No. 600) - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
Section 8-8

PC Block

8-8-1

BEP (Back End Processor) ASSY(FRU No. 600)

Purpose: This is a description on how to remove and replace the Parts in the BEP Assy.
8-8-1-1
Tools
Common pillips screwdrivers
8-8-1-2
Needed Manpower
2 persons, 15 minutes + travel
8-8-1-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
Left Cover, Right Cover and Rear Cover should be removed before proceed. Refer to section 8-5-
1 on page 41, section 8-5-2 on page 43, and section 8-5-3 on page 45.
8-8-1-4

General Procedures of BEP Assy Disassembly.

8.) Open the EMI Cover L Assy (FRU No. 312).
9.) Open the rear cover and EMI Rear Bracket.
10.)Disconnect all the connectors connected With the BEP.
11.)Unscrew eight (8) screws (1-8) and take out the BEP Power plug(9).
12.)Take out the BEP assy to forward direction.
13.)When replace the BEP assy the option dongle should be moved from old BEP to new one. The FRU
BEP doesn't have the option dongle.
8 - 100
7
Figure 8-89 Take out the BEP assy from the console
Section 8-8 - PC Block