7.17 Replacing the frame (from appliance no. 20000)

Tools required:
Spanner set WM 14249,
Allen key 6 mm,
Hexagon socket wrench 8 mm,
Water-pump pliers.
Note:
It is easier to change the frame if you have another person to help you.
1. Open the appliance (see "7.3 Opening the
appliance" on page 29).
2. Remove the compressor (see "7.7 Replacing
the compressor (from appliance no. 15000)"
on page 36, steps
4.
3. Remove the foam section
4. Pull the two appliance connectors
out of the frame.
5. Place one arm around the entire foam
sections.
6. With the other arm, support yourself against
the frame and lift up the foam sections
(including ADS containers, oxygen tank,
cooler, fan and operating panel).
Note:
If two people are working together, one of you
can lift up the foam sections with both arms whilst
the other holds the frame steady.
7. Position the new frame
label) on the castors.
8. Insert the foam sections into the frame.
9. Place the appliance connector
groove of the frame.
10. Align all parts, such as ADS containers and
oxygen tank.
11. Check that the cables and tubes are laid in the
groove on the foam section. If necessary,
press them in.
to
).
8.
.
44
upwards
59
(with appliance
57
into the
59
44
Repair information and repair instructions
59
57
59