Technical Data - Weinmann OXYMAT 3 Servicing And Repair Instructions

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

10. Technical Data

Product category according to
93/42/EEC
Dimensions W x H x D in mm
Weight
Temperature range:
– Operation
– Storage
Power supply
Power output / current consumption
Mains fuse DIN EN 60127-2
Classification acc. to EN 60601-1:
– Protection against electric shock
– Degree of protection against
electric shock
Electromagnetic compatibility:
– Radio interference suppression
– Radio interference resistance
Sound pressure level
O
output
2
(values after 10 min. operation at + 20 °C,
50 % relative humidity and 1013 mbar)
O
output at 2000 m above msl
2
Oxygen outlet pressure
Flow rate (read at top of ball)
O
status indicator OSCI
2
according to DIN EN ISO 8359.51.9
Warning at < 82 VOL % O
Fault warning
Actuating pressure of pressure relief mechanism
Servicing
Materials:
– Case
– Foam sections
Subject to technical change without notice.
2
OXYMAT 3
IIa
400 x 700 x 350
approx. 20 kg
+ 10 to + 40 °C
– 20 to + 70 °C
230 V, 50 Hz
360 W / 1.6 A
T 2.5 A H 250V
Class 2
Type B
EN 55011 (VDE 0875 T.11)
IEC 1000-4 Parts 2-6 and 11
approx. 40 dB(A) (from appliance no. 20000)
95 (+1/–3) Vol % O
1 to 4 l/min
2
90 (± 3) Vol % O
4 to 5 l/min
2
90 ± 3 Vol % O
at 5 l/min
2
0.6 bar = 60 kPa
0.5 to 5.5 l/min
0.2 to 3.0 l/min
Installed
Visual and acoustic
3 bar = 300 kPa
After 1 year / every 5000 operating hours
PP
PP
75
Technical Data

Hide quick links:

Advertisement

Table of Contents
loading