Replacing The Pressure Reducer - Weinmann OXYMAT 3 Servicing And Repair Instructions

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

7.15 Replacing the pressure reducer

Tools required:
Spanner set WM 14249,
Allen key 6 mm,
Open-ended spanner SW 17.
1. Open the appliance (see "7.3 Opening the
appliance" on page 29).
2. Loosen the straps
42
out of the fastener, and place them next to the
appliance.
3. Remove the foam section
4. Carefully raise the operating panel and
detach the connector
5. Carefully pull out the right-hand ADS
container.
6. Pull out the wiring harness for the shutoff valve.
7. Press the ADS container back into the foam
section.
54
Repair information and repair instructions
, without pulling the strap
.
44
for the shutoff valve.
X13
44
X13
42

Hide quick links: