Weinmann OXYMAT 3 Servicing And Repair Instructions page 62

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

Unscrew the piping systems
1. Unscrew the connections to the air inlet and exhaust
hoses and remove the parts. These are:
– Connecting tube, long
– Connecting tube, short
– Push-in connector WM 13146
– Piping system, inlet WM 13270
For concentrators up to appliance no. 19999:
Also remove the exhaust hose and place it ready for
refitting (see point
5.
2.
For concentrators from appliance No. 8000 onwards:
To remove the pressure measurement tube, first turn
the appliance so that the compressor is facing you,
then pull container
67
To ensure safe refitting, note the route followed by
pressure measurement tube
remove the pressure measurement tube
Carefully
from the pressure sensor on the board.
Use a diagonal cutter to cut through the cable
binders, then remove the pressure measurement tube
from the appliance.
Fit valve block to ADS container
1. Place one angle connector
openings on the underside of the valve
block
.
97
2. On one of the ADS containers, replace the
double nipple with the adapter for valve
block
.
104
3. Moisten the O-ring
107
WM 14902.
4. Put the ADS container on the valve block and secure
the connection with the spring clip
62
Repair information and repair instructions
WM 13277
81
WM 13276
80
onwards on page 64).
out of the foam section.
.
63
in each of the
98
with oxygen lubricant
.
105
Exhaust hose
(up to appliance
no.19999
63
67
97
98
104
107
106
105