Weinmann OXYMAT 3 Servicing And Repair Instructions page 64

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

5. Connect up the piping system. To do so, proceed
as follows:
For all concentrators:
Push the piping system
into the connection on the side of the valve block.
For concentrators from appliance No. 20000
onwards:
Connect the exhaust hose to the rear angle
connector of the valve block. For this purpose use
exhaust piping system
For concentrators up to appliance no. 19999:
Fit an angle connector
tube, short
onto the push-in connector of the
80
exhaust hose.
– Wet the rear end of the exhaust hose with
Gaspruf. Push the exhaust hose through the
central foam section onto the exhaust damper.
May sure that the exhaust hose is pushed onto the
damper as far as it will go. While doing so, hold
the exhaust damper in position from the front of
the appliance.
– Push the connecting tube, short
exhaust hose into the rear angle connector of the
valve block.
For all concentrators:
Push the connecting tube
sensor piping system
connector of the valve block.
For concentrators up to appliance no. 7999:
Replace the pressure measurement tube of the
pressure sensor piping system
blanking plug
106
For all concentrators:
Connect the inlet hose to the remaining angle
connector of the pressure sensor piping
system
.
99
64
Repair information and repair instructions
(ADS container) firmly
101
.
100
and a connecting
95
of the
105
of the pressure
105
firmly into the front angle
99
with the
99
.
101
100
Exhaust hose
95
80
From appliance
no. 8000 onwards:
99
inlet hose
80
remaining angle
99
connector