7. Repair information and repair instructions

7.1 General

Repairs to OXYMAT 3 must always be carried out at an ESD workplace!
Please follow the safety instructions on page 4 of
the operating instructions for OXYMAT 3.
Caution!
Do not use liquids or foaming agents to check for
leaks in the region of the solenoid valve 54, as
moisture could penetrate into the solenoid valve,
thereby causing corrosion which could prevent it
functioning properly.
Any handling of the appliance requires a
precise knowledge of and compliance with
the operating instructions and the service and
repair instructions.
Please carry out only the repairs described in
these service and repair instructions. This is the
only way to guarantee perfect functioning of
OXYMAT 3.
Please ensure that your hands and workplace
are clean when carrying out repairs.

7.2 Replacing the castors

1. Position the appliance with the housing lying
on its back.
2. Pull off the defective castors and replace them
with new ones.
28
Repair information and repair instructions
Conduct a functional check after
job (see "5. Functional check" on page 10).
When you replace components or individual
parts, please use only genuine Weinmann
parts.
When ordering a frame 57 or 58 , please also
quote the model, year of construction and
appliance number.
Many parts have to be screwed in using a
torque wrench. To ensure that the quoted
torque is maintained, check regularly that
your torque wrench adheres to the torque
(monitor testing equipment).
Note:
The item numbers quoted in the following text
match the item numbers in the parts list on
page 69 and the overview on page 4.