Weinmann OXYMAT 3 Servicing And Repair Instructions page 47

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

2. Loosen the straps
42
out of the fastener and place them next to the
appliance.
3. Take off the foam section
Replace both ADS containers
4. Undo the Y-connector
sleeve (e.g. using a spanner from the set) and
pull out the angular bush.
5. Tilt the right-hand ADS container
foam section.
6. The ADS container and the magnetic valve
block are connected to one another via an
adapter
. Hold the adapter steady with an
104
open-ended spanner SW19 and unscrew the ADS
container by hand.
7. Now pull the left-hand ADS container out of the
foam section.
8. Release the screwed connection
angle connector
98
Test
1. In order to ascertain whether the ADS
containers really need changing, you should
weigh them.
A container needs changing when it has
absorbed approximately 50 g of moisture, i.e.
when it is 50 g heavier.
Expiry weight = original weight + 50 g.
Please always replace both containers at the
same time.
without pulling the strap
.
44
as follows:
53
by pushing back the
73
out of the
53
incl. the
102
from the double nipple.
44
73
53
104
107
112
102
98
Repair information and repair instructions
42
75
74
72
74
93
106
105
47

Hide quick links:

Advertisement

Table of Contents