Weinmann OXYMAT 3 Servicing And Repair Instructions page 44

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

3. Remove the foam section
Replace both ADS containers
4. Loosen the Y-connector
the sleeve (e.g. using a spanner from the set)
and pull out the angular bush.
5. Pull the ADS container
section.
6. The ADS container
53
are connected by a double nipple. Use an
open-ended spanner to hold the double nipple
firm on the ADS container
magnetic valve
from the double nipple by
54
hand.
7. Lay the open-ended spanner aside and hold
the magnetic valve firmly with your hand.
8. Next, unscrew the ADS container by hand,
including the double nipple.
Test
1. In order to ascertain whether the ADS
container really needs changing, you should
weigh it.
The container needs changing when it has
absorbed approximately 50 g of moisture, i.e.
when it is approximately 50 g heavier.
Expiry weight = Original weight + 50 g.
The two containers should always be changed
at the same time.
44
Repair information and repair instructions
.
44
as follows:
53
by pushing back
73
out of the foam
53
and magnetic valve
54
and unscrew the
53
44
73
53
74
black
white
73
75
72
53
54
75
74
72
74
93
white
black
76
75
73