19. Before assembling the housing, subject the
compressor to a test run:
– Connect the power cord
appliance and to a socket.
– Switch on the appliance.
If no rattling or slipping is audible, the
compressor is correctly installed.
– Switch off the appliance and pull the power
cord
from the socket and from the
10
appliance.

7.6 Replacing the compressor (from appliance no. 13351 to 14999)

Tools required:
Spanner set WM 14249,
Allen key 6 mm,
Water-pump pliers.
1. Open the housing (see "7.3 Opening the
appliance" on page 29).
2. Loosen the straps
42
out of the fastener, and place them next to the
appliance.
3. Remove the front plate
– Tilt the front plate forwards and
– pull it out upwards.
The illustration opposite shows the installation
position of the compressor with its electrical and
pneumatic connections.
to the
10
, without pulling the strap
:
46
20. Close the appliance (see "7.4 Closing the
appliance" on page 30).
21. Perform a functional check (see "5. Functional
check" on page 10).