7.6 Replacing the main circuit board

1. Open the unit (chapter 7.4, page 23).
Remove the main circuit board
1. Pull the plug
and
X200
circuit board.
2. Remove the flat cables from front panel foil
and display:
To do so, pull the lock. Once you have pulled
the lock, you can remove the flat cable.
3. Remove pressure measurement tube
the pressure sensor or pull out the tube adapter
of the pressure measurement tube.
4. der Winkeltülle auf dem Drucksensor.
5. Remove the main circuit board
Inserting the main circuit board
1. Attach tube
to the pressure sensor of a new
45
main circuit board
44
adapter of the pressure measurement tube to
the angular bush on the pressure sensor.
2. Insert the main circuit board.
from the main
X400
from
45
.
44
or connect the tube
SOMNOcomfort 2/SOMNOcomfort 2e
X400
45
Flat cable
SOMNObalance/SOMNObalance e
X200
X400
45
Flat cable
SOMNOcomfort 2/SOMNOcomfort 2e
X400
45
Flat cable
Repair information and repair instructions
X200
44
Lock
44
Lock
X200
44
Lock
25