Weinmann OXYMAT 3 Servicing And Repair Instructions page 48

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

Fit
1. Take up a new ADS container and ensure that
there is a pressure seal
double nipple.
2. Screw the screwed connector onto the double
nipple.
3. Take up the second ADS container and
unscrew the double nipple using the open-
ended spanner SW17.
4. Screw the ADS container onto the adapter of
the magnetic valve block. Hold the adapter
steady with an open-ended spanner SW19 to
do so.
5. Check that the side with the magnetic valve
block coils and the screws for the clip are
flush. If this is not the case, align the ADS
container accordingly, the adapter has a rotating
bearing.
6. Push the Y-connector
the ADS container.
7. Insert the ADS container in the foam section.
Note
Please ensure that there are no cables or tubes
between the ADS container and the foam section.
8. Route cables and/or the tube in the lateral groove
in the foam section.
9. Insert the second ADS container in the foam section
and restore the tube connections.
10. Finally, please check whether the cables and tubes
are correctly routed in the groove in the foam
section.
Important!
The cable for the 230 V voltage supply must be
routed in the left-hand side of the foam section.
11. Before you assemble the OXYMAT 3, check it
for leaks:
– Connect the dosage unit
flowmeter
to the oxygen outlet
16
– Connect the power cord
appliance and to a power socket.
– Switch on the appliance.
– Set a flow of 4 l/min.
– Check for leaks by soaping all connection
points to the ADS containers with a foaming
agent (e.g. Gaspruf).
48
Repair information and repair instructions
on the thread of the
55
into the top angular bush on
73
or the
21
.
5
to the
10
73

Hide quick links: