Test
1.
In order to ascertain whether the ADS
container really needs changing, you should
weigh it.
A container needs changing when it has
absorbed approximately 50 g of moisture,
i.e. when it is approximately 50 g heavier.
Expiry weight = original weight + 50 g.
The two containers should always be changed
at the same time.
Fitting
1. Take a new ADS container and ensure that
there is a pressure seal
double nipple.
2. Screw the screwed connector onto the double
nipple.
3. Take the second ADS container and ensure
that there is a pressure seal
of the double nipple.
4. Screw the double nipple into the magnetic
valve block.
5. Tighten up the double nipple on the magnetic
valve block with the open-ended spanner.
Warning!
Once the pressure seal between the double
nipple and the magnetic valve is squashed, never
screw the magnetic valve back, otherwise the
pressure seal will no longer seal effectively and a
new seal will be required.
6.
Check whether the side with the magnetic
valve block coils and the screws for the clip
are flush. If this is not the case, undo the screw
for the clip a little, align the clip and screw up
tightly.
7. Push the Y-connector
on the ADS container.
8. Insert the ADS container in the foam section.
Note
Please ensure that there are no cables or tubes
between the ADS container and the foam section.
9.
Route cables and/or the tube in the lateral groove
in the foam section.
10. Restore the tube connections.
11. Insert the second ADS container in the foam section
and restore the tube connections.
on the thread of the
55
on the thread
55
into the upper angular bush
73
73
Repair information and repair instructions
51