14. Push clip
onto tube
89
to the compressor inlet and tube
70
compressor outlet. Now attach tube
compressor outlet with the aid of clip
the water-pump pliers.
15. Attach the connector of the compressor to the
power supply socket and lay the cables neatly
into the groove.
16. Check that the compressor is positioned in the
appliance the right way round, i.e. with the
inlet pointing to the right:
– Connect the power cord
appliance and to a socket.
– Switch on the appliance.
– Feel from which side of the compressor air
is being drawn in. This side is the inlet. It
must be on the right.
– Switch off the appliance and pull the power
cord
from the socket and from the
10
appliance.
– If the inlet of the compressor is on the left-
hand side, re-install the compressor.
17. Check that all cables are laid correctly in the
groove.
18. Place the lower edge of the front plate
the frame and tilt it against the foam section
19. Place the straps
over the foam sections and
42
tighten them.
20. Attach the compressor to the chassis:
– Place the appliance on its back.
– Place a washer
90
washer
on the threaded bolt.
91
– Now use the socket wrench to tighten the
hexagonal nuts
92
21. Before assembling the housing, subject the
compressor to a test run:
– Connect the power cord
appliance and to a socket.
– Switch on the appliance.
If no rattling or slipping is audible, the
compressor is correctly installed.
– Switch off the appliance and pull the power
cord
from the socket and from the
10
appliance.
38
Repair information and repair instructions
. Then connect tube
51
to the
51
to the
51
and
89
to the
10
into
46
.
45
and then a spring
hand tight.
to the
10
22. Close the appliance (see "7.4 Closing the
appliance" on page 30).
23. Perform a functional check (see "5. Functional
check" on page 10).
70
89
51
45
46

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents