3. Remove the foam section
4. Loosen the Y-connector
the sleeve (e.g. using a spanner from the set)
and pull out the angular bush. Remove the
outlet piping system WM 13265 from the
appliance.
5. Disconnect plugs
X10
board.
6. Pull the right ADS container
section.
7. The ADS container
53
are connected by a double nipple. Use a 17-
mm open-ended spanner to hold the double
nipple firm on the ADS container
a second open-ended spanner (20 mm) to
unscrew the magnetic valve
double nipple.
Caution!
Be sure to use the 20-mm open-ended
spanner only in the position shown.
Otherwise parts of the magnetic valve could
break off.
8. Next, unscrew the ADS container by hand,
including the double nipple.
9. Repeat points
to
6.
8.
container.
10. Use spanner set WM 14249 to detach the
connecting tube, short
1, 115 long
of the inlet piping system, then
79
remove the right magnetic valve.
11. Use spanner set WM 14249 to detach the
connecting tube, long
1.65 long
and the T and push-in
78
connectors of the inlet piping system. Remove
the left magnetic valve.
.
44
by pushing back
73
and
from the
X11
out of the foam
53
and magnetic valve
54
, and use
53
from the
54
for the left ADS
and hose, PA, 8 x
80
and hose, PA, 8 x
81
44
73
53
53
SW 17
SW 20
54
78
81
79
80
Repair information and repair instructions