Replace Adsorption Containers (Aluminum Magnetic Valve Block) - Weinmann OXYMAT 3 Servicing And Repair Instructions

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

Warning!
Do not use liquids or foaming agents to check for
leaks in the region of the solenoid valve, as
moisture could penetrate into the solenoid valve,
thereby causing corrosion which could prevent it
functioning properly.
12. If you discover a leak, rectify it.
13. Switch off the appliance and pull the power
cord
from the socket and from the
10
appliance.
14. Remove the dosage unit
.
16
15. Put the foam section
16. Place the straps
over the foam sections and
42
tighten them.
17. Close the appliance (see "7.4 Closing the
appliance" on page 30).
18. Perform a final check (see "5. Functional
check" on page 10).
7.13 Replace adsorption containers
(aluminum magnetic valve block)
Tools required:
Spanner set WM 14249
Allen key 6 mm
Open-ended spanner SW17
Water pump pliers
Dismantle
1. Open the appliance (see "7.3 Opening the
appliance" on page 29).
2. Loosen the straps
42
out of the fastener and place next to the
appliance.
or the flowmeter
21
in position.
44
without pulling the strap
Repair information and repair instructions
42
49