Weinmann OXYMAT 3 Servicing And Repair Instructions page 80

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

• Maintenance
Maintenance carried out after 5,000/10,000 etc. operating hours or once every year in accordance with
service instructions
Replacement of suction filter
Replacement of coarse dust filter WM 13631
Maintenance report completed
Faults and possible causes
Repair necessary:
• Final check
Visual inspection (housing and power cord intact, coarse dust filter fitted properly)
Functional test of the control and operating elements
Test of the alarms
O
concentration at 4 l/min: 95 + 1/-3 vol.%
2
If repaired: all functions of the repaired parts tested
• Comments
Findings / maintenance / repair / final repair carried out
Date:
WM-No. 13004
80
Test report
WM 13190
yes
Measured value:
Name:
no
vol.%
Company:
not
OK
OK
not
OK
OK

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents