6. Push plugs
and
X10
block) onto the appropriate connection
points on the board. Carefully locate the
cable in the groove to the side of the right
ADS container.
7. Fit the outlet piping system WM 13265
(non-return valves) to both ADS
containers.
8. Now place the ADS container with
angle connector in the left recess of the
central foam section and push the free
end of the ADS container piping system
onto the angle connector of the left
101
ADS container.
9. Before assembling the OXYMAT 3,
check it for leaks:
– Connect the dosage unit
flowmeter
to the oxygen outlet
16
– Connect the power cord
appliance and to a socket.
– Switch on the appliance.
– Set a flow of 4/min.
– Check for leaks by soaping all
connection points to the ADS
containers with a foaming agent (e.g.
Snoop).
(cables for valve
X11
or
21
.
5
to the
10
Air supply hose
X10, X11
Tubing for outlet
Exhaust air hose
Tubing, exhaust
100
air
Tubing for pressure
99
sensor
Conversion to valve block – view complete
Repair information and repair instructions
Cable guide groove
97
Valve block
Tubing for ADS
101
containers
65