Replace Adsorption Containers (Mac Valves) - Weinmann OXYMAT 3 Servicing And Repair Instructions

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

Caution !
If you do not feel any air flow, remember to
disconnect the appliance from the power before
taking any further action!
11. Switch off the appliance and pull the power
cord
from the socket and from the
10
appliance.
12. Place the lower edge of the front plate
the frame and tilt it against the foam section
13. Place the straps
over the foam sections and
42
tighten them.
14. Align the operating panel.
15. Close the appliance (see "7.4 Closing the
appliance" on page 30).
16. Perform a functional check (see "5. Functional
check" on page 10).

7.11 Replace adsorption containers (Mac valves)

Tools required:
Spanner set WM 14249,
Allen key 6 mm,
Open-ended spanner SW 17.
Dismantle
1. Open the appliance (see "7.3 Opening the
appliance" on page 29).
2. Loosen the straps
42
out of the fastener, and place them next to the
appliance.
into
46
45
, without pulling the strap
.
Repair information and repair instructions
45
46
42
43