Weinmann OXYMAT 3 Servicing And Repair Instructions page 63

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

Prepare second ADS container
1. Replace the pressure seal of the double nipple
with the sealing ring
ring in the screwed connector
2. Screw screwed connector
nipple of the ADS container using a torque
wrench SW 17. Tighten up the screwed
connector
, torque 2.5 Nm.
102
not damaged in the process.
3. Put an angle connector
connector.
Refitting
Note:
Steps 1. to 3. are not necessary for concentrators
up to appliance no. 7999. For these appliances, start with step 4..
1. Carefully push the new pressure measurement tube
(WM 13268, pressure sensor piping system)
99
from the back of the appliance (ADS container)
through the central foam section to the front of the
appliance (compressor).
Take care that the pressure measurement tube is the
right way round. After installation in the appliance,
the end with the T connector and push-in connector
must be at the back of the appliance.
2. At the front of the appliance, lead the pressure
measurement tube
99
and carefully fit the end of the pressure
measurement tube onto the pressure sensor on the
board.
3. Place two new cable binders
pressure measurement tube
exhaust and inlet hoses. Pull the cable binders tight.
Push container
back into the corresponding
67
recess in the central foam section.
4. Place the ADS container with valve block in
the right-hand recess in the central foam
section.
and place the sealing
112
.
102
onto the double
102
Ensure that it is
in the screwed
98
along behind container
around the new
65
, and around the
99
112
102
98
99
67
67
99
67
65
Repair information and repair instructions
63