7.3 Opening the appliance

Tools required:
Spanner set WM 14249,
Allen key 6 mm.
Caution !
If the appliance is standing on a table, lock the castors.
1. Disconnect the plug from the socket and pull
the power cord
from the appliance.
10
3. Remove the stoppers
– Using the disk on the spanner set, rotate the
stopper
through 90° to the left.
32
– Lever it out
downwards
4. Unscrew the two screws
together with the spring washer
To make it easier to remove the screws, set the
spanner in the hexagon and pull out the screw.
5. Remove the handle
1
lying in the appliance.
6. Pull off the housing
35
Note:
If the housing cannot be pulled off, it may be caught
on the service flap. In such cases, release the lock at
the service flap 38, and remove the flap from the
appliance.
:
32
.
and remove them
33
.
34
upwards, or leave it
upwards.
2. Remove the connecting nipple with union nut
or the flowmeter
from the oxygen outlet
16
1
34
33
35
38
Repair information and repair instructions