12. Release the connection between the ADS container
and the magnetic valve by taking the spring clip
out of the adapter for the valve block.
13. Take the ADS container off the magnetic valve.
14. Moisten the O-ring
107
WM 14902
15. Put the ADS container on the new valve block and
secure the connection with the spring clip
Fit magnetic valve block
1. Put the ADS container and valve block into the
right-hand location in the central foam.
2. Prepare the tubing. Proceed as follows to do
so.
– Push the tubing
101
the side connection on the magnetic valve
block.
– Connect the exhaust air hose or on
appliances up to appliance no. 19.999
the connecting pipe for the exhaust air
hose, to the rear angle connector of the
valve block.
– Push the connecting pipe firmly into the front
angle connector of the valve block.
3. Push connectors
X10
valve block) onto the relevant connections of
the circuit board. Guide the cables carefully
into the groove to the side of the right-hand
ADS container.
4. Fit the outlet piping system WM 13265 (non-
return valves) to both ADS containers.
5. Put foam section
44
6. Close the appliance (see "7.4 Closing the
appliance" on page 30).
7. Perform a final check (see "5. Functional
check" on page 10).
68
Repair information and repair instructions
with oxygen lubricant
105
(ADS container) firmly in
and
(cabling for
X11
in position.
105
104
107
.
106
105

Hide quick links: