7.4 Closing the appliance

Tools required:
Spanner set WM 14249,
Allen key 6 mm.
Caution !
If the appliance is standing on a table, please lock the castors.
1. Carefully slide the housing
sections.
2. Position the housing
way that the enclosure wall engages with the
groove all the way round.
3. Insert the handle
and secure it with the
1
screws
(not forgetting the spring washers
33
4. Insert the two stoppers
5. Using the disk on the spanner set, rotate the
stoppers
through 90° to the right.
32
6. Secure the connecting nipple
flowmeter
at the oxygen outlet
16

7.5 Replacing the compressor (up to appliance no. 13350)

Tools required:
Spanner set WM 14249,
Allen key 6 mm.
1. Open the housing (see "7.3 Opening the
appliance" on page 29).
2. Loosen the straps
42
out of the fastener, and place them next to the
appliance.
3. Remove the front plate
– Tilt the front plate forwards and
– pull it out upwards.
30
Repair information and repair instructions