8.3 Closing the device
Tools and equipment required:
•
Phillips screwdriver size 2
1. Carefully put the front and rear case elements
together again.
2. Screw the case together again, making sure
the silicone sealing cord is correctly inserted
and is not jammed or otherwise damaged.
3. Perform a functional check (see "6.2 Perform-
ing the functional check" on page 8).
8.4 Replacing release catch
Tools and equipment required:
•
Phillips screwdriver size 2
•
Screwdriver, size 1 ;
•
Flat or pointed pliers.
1. Open the device (see "8.2 Opening the de-
vice" on page 12).
2. Place the ACCUVAC Basic on its front.
3. Push out retaining pins
and remove them.
63
4. Remove the old or defective release catch
To do so, use a flat / blunt object to press
down snap lock
, which is located below
66
release catch
.
6
5. Take the new release catch
the rear wall of the case.
6. Take O-ring
and locate it in the rear wall of
65
the case between the loops of the release
catch and the rear wall attachment point.
7. Take the retaining pins
the bushing from outside to inside until you
hear them click into place.
8. Close the device (see "8.3 Closing the de-
vice" on page 13).
9. Perform a functional check (see "6.2 Perform-
ing the functional check" on page 8).
from release catch
64
.
63
and insert it in
63
and insert them in
64
52
63
Repairs: Information and Instructions
64
65
66
13