3. Take off the foam section
Replace both ADS containers
4. Undo the Y-connector
sleeve (e.g. using a spanner from the set) and
pull out the angular bush.
5. Take the right-hand ADS container
the foam section.
6. The ADS container and the magnetic valve
block are connected to one another via a
double nipple. Release the double nipple from the
magnetic valve block using an open-ended spanner
SW17. Hold the magnetic valve block steady with
water pump pliers as you do so.
7. Put down the open-ended spanner and the water
pump pliers and hold the magnetic valve block
steady with your hand.
8. Use the other hand to unscrew the ADS container
and the double nipple from the magnetic valve
block.
9. Now take the left-hand ADS container out of
the foam section.
10. Undo the screwed connection incl. angle
connector from the double nipple.
50
Repair information and repair instructions
.
44
as follows:
53
by pushing back the
73
out of
53
44
73
53
112
102
98
75
74
72
74
93
Double nipple
Pressure seal
Aluminum valve
block