Replacing The Circuit Board - Weinmann OXYMAT 3 Servicing And Repair Instructions

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

7.9 Replacing the circuit board

This repair must always be carried out at an ESD workplace.
Tools required:
Spanner set WM 14249,
Allen key 6 mm,
Cross-head screwdriver size1,
Flat nose pliers.
1. Open the appliance (see "7.3 Opening the
appliance" on page 29).
2. Carefully lift the operating panel out of the
foam section.
3. Detach the connectors
circuit board by pressing the lock together and
pulling off the connector.
4. Detach tube
from pressure sensor
63
Caution!
Do not damage pressure sensor when detaching
the tube .
5. The operating panel can now be folded away
upwards.
6. Next, pull connectors
the circuit board.
7. Release the tube
:
84
– Press the lock back using an appropriate
spanner from the set.
– Pull out the tube.
8. Using flat nose pliers,
Off knob
.
2
Caution !
Do not let the pliers slip!
9. To allow you to work more easily, place the
operating panel into the foam section
How to remove the foam section
– Loosen the straps
strap out of the fastener, and place them
next to the appliance.
– Remove the foam section
40
Repair information and repair instructions
and
from the
X1
X2
.
B1
,
and
from
X13
X11
X10
pull off the On/
carefully
.
44
:
44
, without pulling the
42
.
44
84
B1
X13
X11
2
44
44
X1
X2
X10
X1
X2
63
42