Weinmann OXYMAT 3 Servicing And Repair Instructions page 35

Oxygen concentrator
Hide thumbs Also See for OXYMAT 3:
Table of Contents

Advertisement

13. Push clip
onto tube
89
to the compressor inlet and tube
70
compressor outlet. Now attach tube
compressor outlet with the aid of clip
the water-pump pliers.
14. Attach the connector of the compressor to the
power supply socket and lay the cables neatly
into the groove.
15. Check that the compressor is positioned in the
appliance the right way round, i.e. with the
inlet pointing to the right:
– Connect the power cord
appliance and to a socket.
– Switch on the appliance.
– Feel from which side of the compressor air
is being drawn in. This side is the inlet. It
must be on the right.
– Switch off the appliance and pull the power
cord
from the socket and from the
10
appliance.
– If the inlet of the compressor is on the left-
hand side, re-install the compressor.
16. Check that all cables are laid correctly in the
groove.
17. Place the lower edge of the front plate
the frame and tilt it against the foam section
18. Place the straps
over the foam sections and
42
tighten them.
19. Before assembling the housing, subject the
compressor to a test run:
– Connect the power cord
appliance and to a socket.
– Switch on the appliance.
If no rattling or slipping is audible, the
compressor is correctly installed.
– Switch off the appliance and pull the power
cord
from the socket and from the
10
appliance.
20. Close the appliance (see "7.4 Closing the
appliance" on page 30).
21. Perform a functional check (see "5. Functional
check" on page 10).
. Then connect tube
51
to the
51
to the
51
and
89
to the
10
into
46
.
45
to the
10
Repair information and repair instructions
70
89
51
45
46
35

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents