12. Finally, please check whether the cables and
tubes are correctly routed in the groove in the
foam section.
Important!
The cable for the 230 V voltage supply must be
routed in the left-hand side of the foam section.
13. Before you assemble the OXYMAT 3, please
check for leaks:
– connect the dosage unit
to the oxygen outlet
16
– Connect the power cord
appliance and to a power socket.
– Switch on the appliance.
– Set a flow of 4 l/min.
– Check for leaks by soaping all connection
points to the ADS containers with a foaming
agent (e.g. Gaspruf).
Caution!
Do not use liquids or foaming agents to check for
leaks in the region of the solenoid valve, as
moisture could penetrate into the solenoid valve,
thereby causing corrosion which could prevent it
functioning properly.
14. If you discover a leak, rectify it.
15. Switch off the appliance and pull the power
cord
from the socket and from the
10
appliance.
16. Remove the dosage unit
flowmeter
.
16
17. Place the foam section
18. Place the straps
over the foam sections and
42
tighten them.
19. Close the appliance (see "7.4 Closing the
appliance" on page 30).
20. Perform a final check (see "5. Functional
check" on page 10).
52
Repair information and repair instructions
or the flowmeter
21
.
5
to the
10
or the
21
in position.
44