The illustration opposite shows the installation
position of the compressor with its electrical and
pneumatic connections.
4. Disconnect the electrical connection.
5. Pull off the tubes
51
– Take the appropriate spanner from the
spanner set and position it between the
coupling
and the inlet/outlet of the
69
compressor.
– Now pull off the tube.
6. To remove the defective compressor, always
take hold of it at the cylinder head.
Caution !
The cylinder head may still be hot!
Tilt the defective compressor
lift it out.
7. Pull the capacitor
49
section
.
43
Before fitting the new compressor, check the
condition of the vibration dampers
They must be soft, elastic and must not be deformed.
If the vibration dampers are hard and inelastic, they
must be changed.
Then install the new compressor
8. If the capacitor is faulty, disconnect both
spade connectors from the capacitor
9. Connect the spade connectors of the
compressor to the capacitor.
Caution !
Do not short-circuit the capacitor!
The connectors must be on the left and right of the
central bar (see illustration).
and
:
70
forwards and
48
out of the foam
.
60
:
48
.
49
50