Programming the Adjustments

Comman ds used to per for m the adjustmen ts ar e listed in
subsystem of the L4532A/34A Com m a n d Refer en ce. A general sequence
for using these commands for a full adjustmen t is as follows:
CAL:SECURE:STATE OFF,<secure_code> // unsecure for calibration
CAL:STATE ON
CAL:RANGE 0.25 // first range
CAL:RANGE:SOURCE UPP // sets high voltage output
// read the output voltage_hi
CAL:RANGE:SOURCE LOW // sets the low voltage output
// read the low voltage_lo
CAL:RANGE:CHAN? (@1:4), <voltage_hi>,<voltage_lo> // sets all channels
CAL:RANGE 0.5 // do the next range
CAL:RANGE:SOURCE UPP // sets high voltage output
// read the output voltage_hi
CAL:RANGE:SOURCE LOW // sets the low voltage output
// read the low voltage_lo
CAL:RANGE:CHAN? (@1:4), <voltage_hi>,<voltage_lo> // sets all channels
// repeat for ranges 1, 2, 4, 8, 16, 32, 64, 128, 256
// on the 32, 64, 128 and 256 ranges, the high and low input is 0
CAL:STORe // put the new calibration constants into memory
CAL:STATE OFF
CAL:SECURE:STATE ON,<secure_code>
N OT E

Calibration Process

The following gener al procedur e is the r ecommen ded method to complete
a full digitizer calibr ation usin g the Remote Access Pages.
1
2
3
4
5
6
Agilent L4532A/ 34A Service Guide
Note CAL:RANGE:CHAN? and CAL:STOR commands take several seconds to
execute. Be sure to set the remote time-out appropriately
Read
Test Consider ations
Per for m the ver ification tests to character ize the digitizer (incomin g
dat a).
On the Remote Access Page, select the Utilities tab. Click the
Calibration... button to begin the adjustments. Note the calibr ation
count for the digitizer 's maintenan ce r ecords.
Per for m the adjustmen t procedures (
page 68).
Secur e the digitizer against calibr ation .
Note the security code and calibr ation count in the digitizer's
main tenance r ecords.
Calibration and Adjustment
on page 48.