4
Calibration and Adjustment
.
3
4
52
Make a conn ection to the digitizer using the web inter face (see
page
35).
En ter the Remote Control Page and select the
the followin g par ameter s an d click the
Chann els:
Range ±250 mV
Coupling DC
Filter LP_20_MHz
Acquisition :
Sample Rate 1 MSa/ s
Recor ds 1
Samples/ r ecord 100000
Tr ig Holdoff 0 sec
Pr e- tr ig Samples/ r ecord 0
Tr ig delay 0
Tr igger :
Ar m source Immediate
Tr igger source Immediate
L4534A 4-Channel
Configure Digitizer
button.
Apply Configuration
Agilent L4532A/ 34A Service Guide
tab. Set