Dc Accuracy - Agilent Technologies L4532A Series Service Manual

20 msa/ s digitizer
Table of Contents

Advertisement

DC Accuracy

Total Specification ( % of reading + % of range )
Range
250 mV
500 mV
4 V, 64 V
8 V, 128 V
1 V, 2 V, 16 V,
32 V, 256 V
1. 100,000 sample average @ 1 MSa/ s
2. For 1 V range and greater, typical offset with constant temperature is 0.01% of range.
3. Autozero done at room temperature within 24 hours.
Integral nonlinearity
Differential nonlinearity
Input bias current
Agilent L4532A/ 34A Service Guide
23 C 5 C
% of reading
% of range
0.10
0.30
0.10
0.20
0.10
0.30
0.10
0.20
0.10
0.12
1
2,3
T
3 C
T emp coefficient outside 18-28 C
autozero
% of range
% of reading
0.11
0.01
0.06
0.01
0.05
0.01
0.04
0.01
0.04
0.01
5 LSB
1 LSB typical, no missing codes
< 10 nA typical
1
Specifications
/ C
% of range/ C
0.015
0.010
0.015
0.010
0.010
9