Procedure

1
2
3
4
Agilent L4532A/ 34A Service Guide
Select the
Acquire Data
Click the
a
View Measurements...
measur emen ts.
Click
and select Vr ms for each digitizer chann el. Set the Ar ea
b
New
to Win dow and choose AC couplin g. Click the OK button.
Click the
button an d set the St ar ting Offset to 0 and Window
c
Edit..
Size to 40000. Click the OK button.
Leave the Installed Measur emen ts win dow open .
d
On the
tab, click the
Acquire Data
Note the Vr ms measurements for all the chan nels in the Inst alled
a
Measur ements window.
Compar e the measur ed values to the ones shown in the table below.
b
Use the
Configure Digitizer
table below.
Repeat steps 1 and 2 for each range in the table.
If any values ar e out of r an ge, r epair may be r equir ed.
Input
50
T ermination
tab.
button . Delete any inst alled
button .
Run Once
tab to chan ge the Range as in dicated in the
Range
250 mV
500 mV
1 V
2 V
4 V
8V
16 V
Calibration and Adjustment
Error from 0
0.00020 Vrms
0.00024Vrms
0.00032Vrms
0.00050 Vrms
0.0036Vrms
0.0064 Vrms
0.0128 Vrms
4
63