Agilent Technologies L4532A Series Service Manual page 61

20 msa/ s digitizer
Table of Contents

Advertisement

4
Calibration and Adjustment
56
Input Value
Range
-15.2 V
16 V
0 V
+15.2V
-15.2 V
0 V
+15.2 V
-15.2 V
0 V
+15.2 V
-30.4 V
32 V
0 V
+30.4 V
-30.4 V
0 V
+30.4 V
-30.4 V
0 V
+30.4V
-60.8 V
64 V
0 V
+60.8 V
-60.8 V
0 V
+60.8 V
-60.8 V
0 V
+60.8V
Filter
Quick
Check
LP_20_MHz
LP_2_MHz
LP_200_kHz
Q
Q
Q
LP_20_MHz
LP_2_MHz
LP_200_kHz
Q
Q
Q
LP_20_MHz
LP_2_MHz
LP_200_kHz
Q
Q
Q
Agilent L4532A/ 34A Service Guide
Error from Input
0.0344 V
0.0192 V
0.0344 V
0.0344 V
0.0192 V
0.0344 V
0.0344 V
0.0192 V
0.0344 V
0.0688 V
0.0384V
0.0688 V
0.0688 V
0.0384 V
0.0688 V
0.0688 V
0.0384 V
0.0688 V
0.2528 V
0.192 V
0.2528 V
0.2528 V
0.192 V
0.2528 V
0.2528 V
0.192 V
0.2528 V

Advertisement

Table of Contents
loading

This manual is also suitable for:

L4534a series

Table of Contents