1. For this range, set the calibrator output frequency to 300 kHz.
2. (±0.01 dB/°C Temperature Coefficient outside 18-28°C)
to 1 kHz
Agilent L4532A/ 34A Service Guide
Calculate the flatn ess r esult accordin g to the followin g for mula:
Disable the Calibr ator output an d conn ect the Calibr ator to the n ext
chan nel to be verified.
Use the
Configure Digitizer
the Apply Configuration bu tton .
Re- en able the Calibr ator output an d r epeat steps 2 to 8.
Repeat steps 7 to 9 for each chan nel to be ver ified.
Use the
Configure Digitizer
t able below. Set the Calibr ator to output each voltage in the table
below.
Repeat steps 2 to 11 until all the chann els for all the listed r an ges in
the following t able have been ver ified.
If any values ar e out of r an ge, r epair is r equir ed.
Input
165 mVrms
335 mVrms
670 mVrms
1.34 Vrms
2.68 Vrms
43 Vrms
tab to set the Sample Rate to 20 kSa/ s. Click
tab to chan ge the Range as in dicated in the
Range
250 mV
500 mV
1 V
2 V
4 V
1
64
relative
Calibration and Adjustment
Flatness
0.28 dB
0.20 dB
0.20 dB
0.20 dB
0.20 dB
2
0.40 dB
4
61