1
Specifications

Dynamic Characteristics

Measured using a 65536 point FFT.
250 mV
500 mV
1 V
2 V
4 V
8 V
16 V
250 mV
500 mV
1 V
2 V
1. External timebase measurements made with 1 Vpp sinewave with <2 ps RMS jitter.
2. ENOB = (SINAD - 1.76)/6.02
AC flatness (DC-4 MHz)
Crosstalk (Ch to Ch) R
10
1
Input range 980 kHz input (-1 dBFS)
SFDR (dBc)
71
77
81
85
70
70
70
Input range 9.8 MHz input (-1 dBFS)
71
71
69
63
250 mV
500 mV, 1 V, 2 V, 4 V, 8 V, 16 V, 32 V
64 V, 128,V, 256 V
= 50
s
THD (dBc)
SNR (dB)
-79
67
-83
70
-85
73
-82
75
-80
64
-83
65
-81
65
-71
66
-73
68
-68
72
-62
72
0.28 dB relative to 1 kHz
0.20 dB relative to 1 kHz
0.40 dB ( 0.01 dB/ C T emperature
Coefficient outside 18-28 C) relative
to 1 kHz
< -90 dB @ 1 MHz
Agilent L4532A/ 34A Service Guide
2
SINAD (dB)
ENOB
66.7
10.8
69.8
11.3
72.7
11.8
74.2
12.0
63.9
10.3
64.9
10.5
64.9
10.5
64.8
10.5
66.8
10.8
66.5
10.8
61.6
9.9